The Recall Desk

Manufacturer

OPTI Medical Systems, Inc

3 recalls in our database name OPTI Medical Systems, Inc as the manufacturing or recalling firm.

What the numbers show

Total
3
Critical
0
Severe
0
Most recent
2021-08-25

Of the 3 recalls from OPTI Medical Systems, Inc we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–3 of 3

  • HighFDA (Devices)·Z-2310-2021·2021-08-25

    OPTI SARS-CoV-2 RT-PCR Test Kits Recalled for False Positive Results

    OPTI Medical Systems is recalling COVID-19 test kits due to a precipitation issue that may cause false positive results in 5-10% of affected vials. No illnesses have been reported.

    Product
    OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004
    Category
    Medical Device
    Distribution
    10 states
  • ModerateFDA (Devices)·Z-0029-2014·2013-10-30

    OPTI LION E-Plus Cassettes Recalled for Incorrect Barcode Labeling

    OPTI Medical Systems is recalling OPTI LION E-Plus Cassettes because incorrect barcodes may affect performance over time and cause quality control failures.

    Product
    Model BP7507 OPTI LION E-Plus Cassettes 25 Cassettes with Sample Probes, OPTIMedical Cassettes are consumable used in the OPTI LION Electrolyte Analyzer intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and pH in samples of whole blood, se
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2440-2012·2012-10-03

    OPTI Medical Systems Recalls BP7561 B-Lac Cassettes for Altitude Bias

    OPTI Medical Systems is recalling BP7561 B-Lac cassettes because PCO2 readings are biased at high altitudes.

    Product
    BP7561 OPTI CCA-TS B-Lac Cassette (25 per box) Cassettes are consumables used in the OPTI CCA-TS Analyzer intended to be used for in vitro measurements of pH, PCOX2 PO2, lactate (lactic acid), total hemoglobin (tHb), and oxygen saturation (SO2), in heparinized whole blood samp
    Category
    Medical Device
    Distribution
    Distributed nationwide