The Recall Desk

Manufacturer

Noah Medical

4 recalls in our database name Noah Medical as the manufacturing or recalling firm.

What the numbers show

Total
4
Critical
0
Severe
0
Most recent
2025-01-01

Of the 4 recalls from Noah Medical we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–4 of 4

  • HighFDA (Devices)·Z-0742-2025·2025-01-01

    Galaxy System bronchoscope power supply defect may injure lung tissue

    Noah Medical's Galaxy System bronchoscope has defective power supply semiconductors that may fail during procedures, potentially causing pneumothorax or lung tissue injury when the scope is manually removed.

    Product
    Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
    Category
    Medical Device
    Distribution
    15 states
  • HighFDA (Devices)·Z-0693-2025·2024-12-25

    Bronchoscopic system may lose power due to defective component

    A defective power component in Noah Medical's Galaxy System bronchoscope may cause sudden power loss during procedures, potentially leading to lung injury or pneumotharax.

    Product
    Galaxy System (GAL-001)
    Category
    Medical Device
    Distribution
    16 states
  • HighFDA (Devices)·Z-0679-2025·2024-12-18

    Galaxy Bronchoscope Biopsy Tool Insertion Risk at Sharp Bend Angles

    Noah Medical is recalling Galaxy bronchoscope units due to risk that biopsy tools may require excessive force or fail to pass through the scope channel when the bronchoscope is sharply bent, potentially causing tissue injury or pneumothorax.

    Product
    Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1734-2024·2024-05-08

    Galaxy System Wired Controller Recalled for Unintended Scope Motion

    Noah Medical recalls 17 Galaxy System wired controllers (Model GAL-019) due to calibration misalignment that may cause unintended scope motion during bronchoscopic procedures.

    Product
    Galaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide