11 recalls in our database name Neurovision Medical Products Inc as the manufacturing or recalling firm.

What the numbers show

Total
11
Critical
0
Severe
0
Most recent
2025-07-16

Of the 11 recalls from Neurovision Medical Products Inc we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–11 of 11

  • ModerateFDA (Devices)·Z-2100-2025·2025-07-16

    Hummingbird stimulated EMG kit packaging shows the wrong part number

    The Hummingbird Stimulated EMG Kit is being recalled because the product packaging contains the wrong part number and kit contents. 10 units are affected.

    Product
    Brand Name: Hummingbird Product Name: Stimulated EMG Kit with Hummingbird - Model/Catalog Number: REF I-D-BMF90, 1 Single-Use Bipolar Minifork Stimulation Probe, 90mm, and 1 Green Ground Needle ; REF 1-D-BMF90-5, lot number 050925219, expiration 2028-11-30 I-D-BMF90-5 Stim
    Category
    Medical Device
    Distribution
    1 state
  • ModerateFDA (Devices)·Z-2719-2024·2024-09-04

    Cobra EMG ET Tube Recalled Due to Incorrect Size Labeling

    Neurovision Medical Products recalls Cobra 3-Plate EMG ET Tubes due to mislabeling with incorrect tube size on the label. The error could affect proper product selection in clinical procedures.

    Product
    Cobra, 3-Plate EMG ET Tube, 6.0mm, REF: LTE7003PM, not made with natural rubber latex, Rx Only, STERILEEO
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2906-2020·2020-09-02

    Neurovision Cobra x5 EMG Endotracheal Tubes Recalled for Insulation Defect

    Neurovision is recalling Cobra x5 EMG endotracheal tubes because the electrode insulation may soften and damage, potentially stopping EMG signal transmission.

    Product
    Cobra x5 2-Ch EMG ET Tube, 6.0mm, REF: LTE700DCS-5, Rx Only, Sterile EO UDI: B006LTE700DCS52/$$7062618C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventila
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2907-2020·2020-09-02

    Neurovision Cobra x5 EMG ET Tube Recalled for Insulation Softening

    Neurovision is recalling Cobra x5 2-Ch EMG ET Tubes because the electrode insulating material may soften over time, potentially damaging the device and stopping EMG signal transmission.

    Product
    Cobra x5 2-Ch EMG ET Tube, 7.0mm, REF: LTE700DCM-5, Rx Only, Sterile EO UDI: B006LTE700DCM52 - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2908-2020·2020-09-02

    Neurovision Recalls Cobra x5 2-Ch EMG ET Tubes Due to Insulation Defect

    Neurovision Medical Products Inc is recalling 1,058 Cobra x5 2-Ch EMG ET Tubes because the electrode insulating material may soften, potentially stopping EMG signal transmission.

    Product
    Cobra x5 2-Ch EMG ET Tube, 8.0mm, REF: LTE700DCL-5, Rx Only, Sterile EO UDI: B006LTE700DCL52 - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2903-2020·2020-09-02

    Neurovision Cobra EMG ET Tube Recalled for Insulating Material Softening

    Neurovision is recalling Cobra x5 EMG ET tubes because the insulating material may soften and damage, potentially stopping EMG signal transmission.

    Product
    Cobra x5 1-CH EMG ET Tube, 6.0mm, REF: LTE700S-5, Rx Only, Sterile EO UDI: B006:TE700S52$$7050918B - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2905-2020·2020-09-02

    Neurovision Cobra EMG Endotracheal Tubes Recalled for Insulation Softening

    Neurovision is recalling Cobra x5 1-Ch EMG ET Tubes because the electrode insulating material may soften over time, potentially damaging the device and stopping EMG signal transmission.

    Product
    Cobra x5 1-Ch EMG ET Tube, 8.0mm, REF: LTE700L-5, Rx Only, Sterile EO UDI: B006LTE700L52$$7061318B - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2900-2020·2020-09-02

    Neurovision Nerveana x5 RLN Monitoring Kit Recalled for Insulation Defect

    Neurovision is recalling Nerveana x5 RLN Monitoring Kits because the electrode insulating material may soften, potentially stopping EMG signal transmission and rendering the device unusable.

    Product
    Nerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMG ET Tube, 6.0mm REF NVTKIT-C6-5, Rx Only, Sterile, UDI: B006NVTKITC652/$$7061918A - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2902-2020·2020-09-02

    Neurovision Nerveana x5 RLN Monitoring Kit Recalled for Insulation Softening

    Neurovision is recalling Nerveana x5 RLN Monitoring Kits because the electrode insulating material may soften, potentially damaging the device and stopping EMG signal transmission.

    Product
    Nerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMG ET Tube, 8.0 mm, REF NVTKIT-C8-5, Rx Only, Sterile UDIB006NVTKITC852/$$7062118C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2901-2020·2020-09-02

    Neurovision Nerveana x5 RLN Monitoring Kit Recalled for Insulation Defect

    Neurovision Medical Products Inc. is recalling Nerveana x5 RLN Monitoring Kits because the electrode insulating material may soften, potentially stopping EMG signal transmission and rendering the device unusable.

    Product
    Nerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMT ET Tube, 7.0mm REF: NVTKIT-C7-5, RxOnly, Sterile, UDI: B006NvtKITc752/$$7062818C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an ope
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2904-2020·2020-09-02

    Neurovision Cobra EMG ET Tube Recalled for Insulating Material Softening

    Neurovision Medical Products Inc. is recalling Cobra x5 1-Ch EMG ET Tubes because the electrode insulating material may soften, potentially damaging the device and stopping signal transmission.

    Product
    Cobra x5 1-Ch EMG ET Tube, 7.0mm, REF LTE700M-5, Rx Only, Sterile EO CE UDI: B006LTE700M52/$$7061418C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation
    Category
    Medical Device
    Distribution
    Distributed nationwide