The Recall Desk

Manufacturer

Mercury Enterprises, Inc. dba Mercury Medical

3 recalls in our database name Mercury Enterprises, Inc. dba Mercury Medical as the manufacturing or recalling firm.

What the numbers show

Total
3
Critical
0
Severe
2
Most recent
2025-01-22

Of the 3 recalls from Mercury Enterprises, Inc. dba Mercury Medical we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2 (67%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–3 of 3

  • SevereFDA (Devices)·Z-0821-2025·2025-01-22

    Neo-Tee Resuscitator Spring Defect May Impair Pediatric Ventilation

    Mercury Medical is recalling the Neo-Tee T-Piece Resuscitator due to an undersized spring in the circuit flow controller that may prevent reaching proper pressure ranges, potentially impairing emergency ventilation.

    Product
    Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, 1050841, 1050842
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0004-2025·2024-10-09

    Neo-Tee T-Piece Resuscitator circuit flow controller may come apart

    Mercury Medical is recalling approximately 1,300 Neo-Tee T-Piece Resuscitators due to a manufacturing defect that may cause the circuit flow controller to come apart, potentially compromising emergency respiratory support.

    Product
    Neo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.
    Category
    Medical Device
    Distribution
    5 states
  • HighFDA (Devices)·Z-0639-2016·2016-01-20

    Mercury Medical Recalls Neo-Tee T-Piece Resuscitators Due to Occlusion Risk

    Mercury Medical is recalling Neo-Tee T-Piece Resuscitators and circuits because incorrect use can cause complete circuit occlusion, leading to failure to ventilate the patient.

    Product
    Mercury Medical Neo-Tee T-Piece Resuscitator and T-Piece Circuit. The Neo-Tee T-Piece Resuscitator is a gas powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use
    Category
    Medical Device
    Distribution
    Distributed nationwide