9 recalls in our database name Med Tec Inc as the manufacturing or recalling firm.

What the numbers show

Total
9
Critical
0
Severe
0
Most recent
2020-06-10

Of the 9 recalls from Med Tec Inc we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–9 of 9

  • HighFDA (Devices)·Z-2243-2020·2020-06-10

    Med Tec Type S Extension Recalled for Potential Disengagement Risk

    Med Tec Inc is recalling Type S Extensions used for patient positioning due to a risk of disengagement during repositioning, which could cause a patient fall.

    Product
    Type S Extension For use with Varian, CIVCO, Prescription Only - Product Usage: Device is intended for patient positioning.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1837-2019·2019-06-26

    CIVCO Solstice SRS Immobilization System Recalled for Potential Movement

    Med Tec Inc is recalling 28 units of the CIVCO Solstice SRS Immobilization System because the device may move during patient setup or treatment.

    Product
    CIVCO Solstice(TM) SRS Immobilization System, Part Numbers: a) CHS03, b) CHS04 - Product Usage: The Tilting Head Fixation (Solstice) device is comprised of a baseplate and bowl with a locking lever. The bowl is permanently attached to the baseplate and the locking level allows t
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0671-2018·2018-02-28

    Med Tec Protura Software Recalled for Pedestal Coordinate Errors

    Med Tec Inc is recalling Protura software versions that may fail to update pedestal location data, potentially affecting treatment table movements.

    Product
    Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interfac
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-3082-2017·2017-09-06

    Med Tec Rigid Arm Supports Recalled for Height Dimensional Variance

    Med Tec Inc is recalling 101 Rigid Arm Supports used in Posiboard-2 and Posirest systems due to a potential 17mm height difference.

    Product
    CIVCO Arm Support, REF 106015, distributed as: (a) Rigid Arm Support, Standard, Set Red (PN: 106015) (b) Rigid Adj. Arm Support, Set (PN: 106020) (c) PSR2-CR-DLR-2P Posirest-2 (PN: 106521) (d) PB2-BLR Posiboard-2 (PN: 109030) The Rigid Arm Supports are utilized on the Po
    Category
    Medical Device
    Distribution
    3 states
  • ModerateFDA (Devices)·Z-3083-2017·2017-09-06

    Med Tec Recalls CIVCO Rigid Arm Supports for Height Discrepancy

    Med Tec Inc is recalling CIVCO Rigid Arm Supports used in Posiboard-2 and Posirest assemblies due to a potential 17mm height difference.

    Product
    CIVCO Arm Support, REF 106047, distributed as: (a) Rigid Arm Support, Large, Set, Red (PN: 106047) (b) PB2-BDLRL (PN: 109040) (c) Rigid Adj. Arm Support, Large, Set (PN: 106048) (d) PSR2-CF-DLRL-2P Posirest-2 (PN: 106522) The Rigid Arm Supports are utilized on the Posibboa
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-1992-2016·2016-06-22

    Med Tec Recalls Protura Robotic Couch System Software Due to Positioning Issues

    Med Tec Inc is recalling specific versions of Protura Robotic Couch System Software because non-sub-millimeter positioning is inconsistent with settings in split shift mode.

    Product
    Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT6XSM1.7.1), CIVCO, Orange City, IA 51041
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-2763-2015·2015-09-23

    Med Tec Recalls CIVCO Shoulder Retractors Due to Excessive Strap Length

    Med Tec Inc is recalling CIVCO Shoulder Retractors because the wrist straps used for patient positioning are too long.

    Product
    CIVCO Shoulder Retractors, Reusable non-sterile hand grip, REF 302070, packaged 1 set per box, RX. For use in patient positioning. The grips are used for pulling down the shoulders during radiotherapy simulation and treatment sessions where opposed lateral fields are used in
    Category
    Medical Device
    Distribution
    3 states
  • LowFDA (Devices)·Z-0990-2015·2015-01-21

    Med Tec Recalls Short Wool Wrist Loops for Shoulder Retractors

    Med Tec Inc is recalling specific wool wrist loops for shoulder retractors because they were manufactured 7cm shorter than specifications.

    Product
    Civco, MTSR03 Shoulder Retractor with Wrist Loops Kit with Base, plus two pairs of wrist loops (nylon and sherpa), Non-sterile re-usable wool wrist strap, REF: 20SR03, Rx The devise is intended to pull patients shoulders down for clearance when treating lateral fields to the
    Category
    Medical Device
    Distribution
    4 states
  • LowFDA (Devices)·Z-0989-2015·2015-01-21

    Med Tec Recalls Short Wool Wrist Loops for Shoulder Retractors

    Med Tec Inc is recalling 11 units of a wool wrist loop component for shoulder retractors because the items were manufactured 7cm shorter than specifications.

    Product
    Civco, MTSR02 Shoulder Retractor with Sherpa Wrist Loops Kit with Base, Non-sterile re-usable wool wrist strap, REF: 20SR02SUB2, Rx Only, The devise is intended to pull patients shoulders down for clearance when treating lateral fields to the neck.
    Category
    Medical Device
    Distribution
    4 states