6 recalls in our database name Med-Mizer, Inc. as the manufacturing or recalling firm.

What the numbers show

Total
6
Critical
0
Severe
3
Most recent
2023-03-08

Of the 6 recalls from Med-Mizer, Inc. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 3 (50%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–6 of 6

  • HighFDA (Devices)·Z-1205-2023·2023-03-08

    Sit to Stand STS500 Patient Lift Recalled Due to Loose Boom Fasteners

    Med-Mizer is recalling 37 units of the Sit to Stand STS500 Patient Lift due to risk of boom pivot failure caused by a loose securing nut and unsecured bolt on the boom flange.

    Product
    Sit to Stand STS500 Patient Lift
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1204-2023·2023-03-08

    Med-Riser MR600 Patient Lift Boom Pivot Failure Risk Recall

    Med-Mizer, Inc. is recalling the Med-Riser MR600 Patient Lift due to risk that the boom pivot may fail if the securing nut loosens and the bolt becomes unsecured. Contact the manufacturer for instructions.

    Product
    Med-Riser MR600 Patient Lift
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-3183-2018·2018-09-26

    Med-Mizer Recalls PR1000 Pivot Rehab Beds Due to Cable Pinching Risk

    Med-Mizer is recalling 230 PR1000 Pivot Rehab Beds because wires may become pinched during extreme left pivoting, requiring patient transfer until repaired.

    Product
    PR1000 Pivot Rehab Bed AC powered adjustable bed.
    Category
    Medical Device
    Distribution
    28 states
  • SevereFDA (Devices)·Z-1542-2014·2014-05-14

    Med-Mizer RetractaBed Recall for Insufficient Welding on Foot Cross Tube

    Med-Mizer is recalling RetractaBed Clinical Contour adjustable beds due to insufficient welding on the foot cross tube, which may cause bed collapse or rollers to detach.

    Product
    RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 799, 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1541-2014·2014-05-14

    Med-Mizer Retractabed Recall for Insufficient Welding on Foot Cross Tube

    Med-Mizer is recalling 191 Retractabed Clinical Contour adjustable beds manufactured before June 7, 2008, due to insufficient welding that could cause bed collapse.

    Product
    RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors); 763 (3 motors); 763 A ( 3 motors, autocontour) 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1543-2014·2014-05-14

    Med-Mizer RetractaBed Recall Due to Insufficient Welding on Foot Cross Tube

    Med-Mizer is recalling 2,980 RetractaBed Clinical Contour adjustable beds due to insufficient welding on the foot cross tube, which may cause bed collapse.

    Product
    RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 801, 802, 803, 803A: 80 " bed with 1, 2, or 3, motors and autocontour 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
    Category
    Medical Device
    Distribution
    Distributed nationwide