GE Healthcare (China) Co., Ltd.

24 recalls in our database name GE Healthcare (China) Co., Ltd. as the manufacturing or recalling firm.

What the numbers show

Total
24
Critical
0
Severe
0
Most recent
2025-03-05

Of the 24 recalls from GE Healthcare (China) Co., Ltd. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–24 of 24

  • HighFDA (Devices)·Z-1239-2025·2025-03-05

    MRI Systems Allow Unsafe Scan Resumption Exceeding SAR Safety Limits

    GE Healthcare MRI systems may incorrectly resume scans after power loss when tissue heating rates exceed safe limits, risking excessive heating if patients contact conductive materials or have implants.

    Product
    SIGNA Creator, SIGNA Explorer, NMRI systems
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1241-2025·2025-03-05

    GE Healthcare MRI system recalled for RF energy safety control failure

    GE Healthcare's Optima MR360 MRI systems may allow unsafe RF energy exposure if a scan resumes after power loss. The user interface can incorrectly enable scanning when tissue-heating levels exceed safe limits.

    Product
    Optima MR360, NMRI system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1243-2025·2025-03-05

    GE SIGNA MR380 NMRI System SAR Limit Exceeded During Scan Resumption

    GE SIGNA MR380 NMRI systems may allow scans to resume when the 6-minute average SAR (tissue heating rate) exceeds safe limits, potentially causing excessive tissue heating especially in patients with implants or conductive contact.

    Product
    SIGNA MR380, NMRI system, Non-US Only
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1242-2025·2025-03-05

    MRI Systems May Resume Scans When SAR Exceeds Safe Limits

    Certain GE Healthcare MRI systems may incorrectly enable scan resumption when radiofrequency energy absorption exceeds FDA-set limits, potentially causing tissue heating. Risk is higher for patients with implants or touching conductive materials.

    Product
    SIGNA MR355, SIGNA MR360, NMRI system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1240-2025·2025-03-05

    MRI System May Resume Scanning Above Safe Thermal Limits

    GE Healthcare's Brivo MR355 MRI system can incorrectly resume scans when tissue heating limits are exceeded, potentially causing thermal burns if patients contact conductive materials. The recall affects 22 systems distributed nationwide and internationally.

    Product
    Brivo MR355, NMRI system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0601-2025·2024-12-11

    GE Healthcare Optima 520 CT systems recalled for image rotation

    GE Healthcare recalled 11 Optima 520 CT systems that produce progressively rotated scan images in helical, cine, and cardiac modes, potentially affecting diagnostic accuracy.

    Product
    GE Healthcare Optima 520, Catalog number 5439126; X-ray/computed tomography systems
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0610-2025·2024-12-11

    GE Healthcare Revolution Maxima CT Systems Recalled for Progressive Image Rotation

    GE Healthcare recalled Revolution Maxima and related CT imaging systems that can produce progressively rotated images during helical, cine, and cardiac scans. Approximately 72 units worldwide are affected; no patient injuries reported.

    Product
    GE Healthcare Revolution Maxima, Model Numbers: 1) 6665000-2, 2) 6670000-2, 3) 6670000-200, 4) 6670000-6; X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0603-2025·2024-12-11

    GE Healthcare Revolution ACE CT Systems Image Rotation Issue

    GE Healthcare's Revolution ACE CT systems can produce progressively rotated images during certain scan modes. Maximum rotation reaches 56 degrees, potentially affecting diagnostic interpretation.

    Product
    GE Healthcare Revolution ACE, Model number 6670000-3; X-ray/computed tomography systems
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0600-2025·2024-12-11

    GE Healthcare CT Systems May Display Rotated Images in Diagnostic Scans

    Certain GE Healthcare CT systems may produce progressively rotated images during diagnostic scans. If undetected and a rescan is needed, the same rotation recurs.

    Product
    GE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as:
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0605-2025·2024-12-11

    GE Healthcare Revolution Ascend CT System Recalled for Image Rotation

    GE Healthcare Revolution Ascend CT systems can produce progressively rotated images during helical, cine, and cardiac scans. Images after the first may be rotated up to 56 degrees, potentially affecting diagnostic accuracy.

    Product
    GE Healthcare Revolution Ascend, Model number 6969000-100; X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0613-2025·2024-12-11

    GE Healthcare CT Operator Console Images May Rotate During Scans

    GE Healthcare CT systems may produce progressively rotated images during helical, cine, and cardiac scans. Images can rotate up to 56 degrees, potentially requiring patient rescans.

    Product
    GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B79632DA: EU: RT Refresh Package, Non-EU: Discovery RT, Optima CT580, or Discovery CT590 RT; X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0611-2025·2024-12-11

    GE Healthcare CT System May Produce Rotated Diagnostic Images

    GE Healthcare's Revolution Maxima M and related CT systems may produce progressively rotated images during scans, potentially compromising diagnostic accuracy. Rotation increases with each image, reaching up to 56 degrees.

    Product
    GE Healthcare Revolution Maxima M, Model Number, 6670000-2; X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0615-2025·2024-12-11

    GE Healthcare CT Operator Console Update: Images May Rotate Progressively

    GE Healthcare operator console upgrades for certain CT systems can cause progressive image rotation during scans. Images may rotate up to 56 degrees per exam, potentially affecting diagnostic accuracy.

    Product
    GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80522DA: China Only: Revolution Discovery CT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0604-2025·2024-12-11

    CT imaging systems may display progressively rotated images in certain scan modes

    Certain GE Healthcare CT systems may display progressively rotated images, particularly during helical, cine, or cardiac scans. The rotation increases with each image, potentially reaching 56 degrees maximum.

    Product
    GE Healthcare Revolution Advance, Model number 5849504; X-ray/computed tomography systems
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0607-2025·2024-12-11

    GE Healthcare CT Imaging System May Produce Rotated Medical Images

    GE Healthcare CT systems may produce progressively rotated images during helical, cine, and cardiac scans. The issue affects 4 units worldwide and can cause rotations up to 56 degrees.

    Product
    GE Healthcare Revolution Eagle, Model Number 5849504-3, X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0612-2025·2024-12-11

    GE Healthcare CT systems produce progressively rotated images in scan modes

    Certain GE Healthcare CT systems can produce progressively rotated images during helical and cardiac scans. Images may rotate up to 56 degrees, potentially affecting diagnosis if the rotation is not detected.

    Product
    GE Healthcare Revolution Maxima Select, Model Number, 6670000-55; X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0606-2025·2024-12-11

    CT imaging system issue causes progressive image rotation during scans

    GE Healthcare CT systems have a technical issue causing progressive image rotation during helical and cardiac scans. No injuries reported, but unnoticed rotation could affect diagnosis or require rescans.

    Product
    GE Healthcare Revolution Ascend Elite, Model Number 6967000-100; X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0608-2025·2024-12-11

    GE Healthcare CT Systems Recalled Due to Progressive Image Rotation

    GE Healthcare CT systems may display progressively rotated images, particularly in helical, cine, and cardiac scans. Rotations up to 56 degrees may affect image interpretation and could necessitate additional patient scanning.

    Product
    GE Healthcare Revolution EVO, Model Number 5454001-260; X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-0609-2025·2024-12-11

    GE Healthcare Revolution Frontier CT systems may produce rotated images during scans

    GE Healthcare Revolution Frontier CT systems can produce progressively rotated images during helical, cine, and cardiac scans. The rotation can reach up to 56 degrees and may require repeat scans.

    Product
    GE Healthcare Revolution Frontier, Model Numbers: 1) 5232085-213, 2) 5873009; X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-0614-2025·2024-12-11

    GE Healthcare CT Imaging Systems Recalled for Progressive Image Rotation

    GE Healthcare is recalling operator console upgrades for certain CT systems due to a software defect causing progressive image rotation during scans. Rotations can reach up to 56 degrees and may require patient rescanning.

    Product
    GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80212DA: EU: Operator Console (for BrightSpeed, Elite/BrightSpeed Elite Select), Non-EU: BrightSpeed Elite, BrightSpeed Elite Select; X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-0616-2025·2024-12-11

    GE Healthcare CT Scanner Console Images May Rotate After First Scan

    Certain GE Healthcare CT scanner systems may produce progressively rotated images after the first scan, with rotation up to 56 degrees. While anatomical data remains intact, misaligned images could necessitate patient rescans.

    Product
    GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80662DA: China Only: Discovery CT750HD
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-0602-2025·2024-12-11

    CT Image Rotation Issue in GE Healthcare Optima 680 Systems

    GE Healthcare Optima 680 Expert CT systems may produce progressively rotated images during helical and cardiac scans. Affected images may require rescanning for accurate clinical interpretation.

    Product
    GE Healthcare Optima 680 Expert, Model number 6670000; X-ray/computed tomography system
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-0550-2025·2024-12-04

    GE Proteus XR/A radiographic system Wall Stand cable maintenance recall

    GE Healthcare is recalling certain Proteus XR/A radiographic systems because steel cables in the Wall Stand component have not been replaced at manufacturer-specified maintenance intervals.

    Product
    GE Proteus XR/A radiographic system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2733-2024·2024-09-04

    GE Healthcare Proteus XR/a X-Ray System Missing De-Installation Manual

    GE Healthcare's Proteus XR/a stationary X-ray system lacks a required de-installation manual, creating risk of personnel injury during device removal.

    Product
    GE Healthcare Proteus XR/a, stationary X-Ray System, Model/Catalog Numbers: 1) 2259973; 2) 2259988; 3) 2266999; 4) 2373519; 5) 2379827; 6) 5729352; 7) 2259973-1; 8) 2259973-2; 9) 2259973-3; 10) 2259973-4; 11) Not Available
    Category
    Medical Device
    Distribution
    0 states