The Recall Desk

Manufacturer

Datex-Ohmeda, Inc.

30 recalls in our database name Datex-Ohmeda, Inc. as the manufacturing or recalling firm.

What the numbers show

Total
30
Critical
1
Severe
21
Most recent
2026-09-23

Of the 30 recalls from Datex-Ohmeda, Inc. we have scored against our rubric, 1 (3%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 21 (70%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–25 of 30

  • HighFDA (Devices)·Z-3192-2026·2026-09-23

    Datex-Ohmeda Et Control Software Recalled for Anesthetic Delivery Interruption

    Datex-Ohmeda is recalling 410 Et Control software units for anesthesia gas machines that may interrupt anesthetic delivery, requiring clinician intervention to prevent decreased patient effect.

    Product
    Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3189-2026·2026-09-23

    Datex-Ohmeda Aisys Anesthesia Systems Recalled for Gas Mixer Defect

    Datex-Ohmeda is recalling Aisys anesthesia systems because electronic gas mixer assemblies may fail, interrupting anesthetic delivery and requiring clinician intervention.

    Product
    Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3190-2026·2026-09-23

    Datex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect

    Datex-Ohmeda is recalling 140 Aisys CS2 anesthesia systems because electronic gas mixer assemblies may fail, interrupting anesthetic delivery and requiring clinician intervention.

    Product
    Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3191-2026·2026-09-23

    Datex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect

    Datex-Ohmeda is recalling seven Aisys CS2 anesthesia systems because a gas mixer defect can interrupt anesthetic delivery, requiring clinician intervention to prevent decreased patient effect.

    Product
    Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2357-2026·2026-06-24

    Giraffe infant resuscitation system may deliver wrong oxygen concentration

    Giraffe Stand-alone Infant Resuscitation Systems are being recalled because a loosening blender knob shaft can cause delivery of an oxygen concentration that does not match the setting.

    Product
    Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001, 2082663-002, M1241424 Product impacted only if service blender M1091607-R was installed during the last Preventative Maintenance
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-2359-2026·2026-06-24

    Panda freestanding warmer blender knobs may deliver wrong oxygen concentration

    Panda Freestanding Warmers are being recalled because a loosening blender knob shaft can cause delivery of an oxygen concentration that does not match the setting.

    Product
    Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-2358-2026·2026-06-24

    Giraffe and Panda warmer blender knobs may deliver wrong oxygen concentration

    Giraffe and Panda Warmers with Integrated Resuscitation System are being recalled because a loosening blender knob shaft can cause delivery of an unintended oxygen concentration.

    Product
    Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0141-2025·2024-11-06

    GE Healthcare Infant Incubator Recall: Airborne Formaldehyde Off-Gassing

    GE Healthcare Giraffe Incubator Carestation CS1 units may release airborne formaldehyde residue during initial use. Manufacturer notified customers in June 2023 to run incubators for one week to off-gas any residual formaldehyde, which can cause eye irritation.

    Product
    GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3) 2082844-002-01058424; 4) 2082844-002-01069017; 5) 2082844-002-01089666; 6) 2082844-002-01090474; 7) 2082844-002-01091269; 8)…
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0140-2025·2024-11-06

    GE Healthcare Giraffe infant incubator recalled for formaldehyde exposure

    GE Healthcare Giraffe Omnibed infant incubators may release airborne formaldehyde when newly manufactured. Customers were notified in June 2023 to run units for one week to reduce potential off-gassing.

    Product
    GE Healthcare Giraffe Omnibed Carestation CS1, combination infant incubator and infant warmer. Model Numbers: 1) 2082844-001-01040418; 2) 2082844-001-01046784; 3) 2082844-001-01052324; 4) 2082844-001-01052572; 5) 2082844-001-01055337; 6) 2082844-001-01065693; 7)…
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0076-2025·2024-10-30

    Aespire 7900 Anesthesia Gas Machine Recalled Over Breathing Circuit Misconnection Risk

    Datex-Ohmeda has recalled 12,036 Aespire 7900 anesthesia gas machines worldwide due to potential risk of patient breathing circuit being mistakenly connected to the Auxiliary Common Gas Outlet port.

    Product
    Aespire 7900. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0074-2025·2024-10-30

    Aespire 7100 Anesthesia Machines Recalled for Breathing Circuit Misconnection

    Datex-Ohmeda is recalling 20,097 Aespire 7100 series anesthesia machines because a patient breathing circuit can be mistakenly connected to the wrong outlet (ACGO port) instead of the intended connection point.

    Product
    Aespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0075-2025·2024-10-30

    Aespire 100 anesthesia machine breathing circuit misconnection FDA recall

    Datex-Ohmeda Aespire 100 anesthesia machines are recalled due to potential misconnection of breathing circuit limbs to the Auxiliary Common Gas Outlet port, which could impair anesthesia or ventilation delivery.

    Product
    Aespire 100. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0081-2025·2024-10-30

    Anesthesia Machine Recall Due to Patient Breathing Circuit Misconnection Risk

    Datex-Ohmeda Aisys anesthesia machines may allow patient breathing circuits to be incorrectly connected to the Auxiliary Common Gas Outlet, potentially affecting ventilation during anesthesia. Approximately 4,790 units worldwide are affected.

    Product
    Aisys. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0077-2025·2024-10-30

    Datex-Ohmeda Aespire View Anesthesia Machines Recalled for Connection Error Risk

    Datex-Ohmeda is recalling 10,338 Aespire View anesthesia machines due to a potential issue where a patient breathing circuit limb may be mistakenly connected to the Auxiliary Common Gas Outlet port.

    Product
    Aespire View. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0080-2025·2024-10-30

    Anesthesia gas machines pose risk of breathing circuit misconnection

    Avance CS2 and Avance CS2 Pro anesthesia machines may permit breathing circuit misconnection to the wrong gas outlet. Affected devices have the AGCO option and are distributed worldwide.

    Product
    Avance CS2 and Avance CS2 Pro. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0083-2025·2024-10-30

    Aisys CS2 Anesthesia Machine Breathing Circuit Misconnection Hazard

    Datex-Ohmeda's Aisys CS2 anesthesia machine may pose a risk if a patient breathing circuit limb is mistakenly connected to the Auxiliary Common Gas Outlet port. Ninety-two units worldwide are affected in this FDA Class I recall.

    Product
    Aisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0079-2025·2024-10-30

    Amingo Anesthesia System: Risk of Breathing Circuit Misconnection to Auxiliary Outlet

    Datex-Ohmeda Amingo anesthesia systems may allow a patient breathing circuit to be mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) instead of the proper outlet. The potential misconnection affects all 77 units worldwide.

    Product
    Amingo. OUS only.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0078-2025·2024-10-30

    Anesthesia Gas Machine Recall: Breathing Circuit Misconnection Risk

    Datex-Ohmeda is recalling Avance anesthesia gas machines worldwide due to a potential hazard if a patient's breathing circuit is incorrectly connected to the Auxiliary Common Gas Outlet port during setup.

    Product
    Avance. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0082-2025·2024-10-30

    Aisys CS2 Anesthesia Machine Recall: Breathing Circuit Misconnection Risk

    Datex-Ohmeda is recalling approximately 15,739 Aisys CS2 anesthesia machines worldwide. A breathing circuit limb may be mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port instead of the intended connection point.

    Product
    Aisys CS2. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0084-2025·2024-10-30

    Datex-Ohmeda Auxiliary Common Gas Outlet Conversion Kits Recalled

    Datex-Ohmeda is recalling 649 Auxiliary Common Gas Outlet (ACGO) conversion kits due to a potential issue if a patient breathing circuit limb is mistakenly connected to the ACGO port.

    Product
    Auxiliary Common Gas Outlet (ACGO) conversion/install kits sold under product names CONV KIT SCGO TO ACGO and ACGO ASSY FIELD INSTL. These kits are not affected specifically, however are included due to being purchased with or later used to convert to ACGO.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0048-2025·2024-10-23

    GE HealthCare Recalls Giraffe OmniBed Carestations Due to Door Screw Issue

    GE HealthCare is recalling certain Giraffe OmniBed Carestations because a screw securing the heater doors may not be torqued to specification, potentially causing the doors to become loose.

    Product
    GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0047-2025·2024-10-23

    GE HealthCare Giraffe OmniBed neonatal incubator door screw may loosen

    Certain GE HealthCare Giraffe OmniBed and Carestation neonatal incubators may have door-securing screws that were not torqued to specification, potentially causing the doors to become loose. These units are distributed worldwide.

    Product
    GE Heathcare Giraffe OmniBed; neo natal incubator
    Category
    Medical Device
    Distribution
    0 states
  • CriticalFDA (Devices)·Z-0776-2024·2024-02-07

    Medical Device Compressor Kits Recalled for Elevated Formaldehyde Levels

    GE HealthCare is recalling EvAir compressor kits for CARESCAPE and Engström ventilators due to elevated formaldehyde levels found in preliminary testing. The recall affects 1,294 units distributed worldwide.

    Product
    EvAir CPRSR Kit Upgrade: a) DISS/CEE 7/7 230V 50/60 HZ, Model Number M1230849; b) DISS/NEMA 5-15 115V 60 HZ, Model Number M1230847; c) NIST/BS 1363 230V 50/60 HZ, Model Number M1230849; d) NIST/BS 546 230V 50/60 HZ, Model Number M1230849; e) NIST/CEE 7/7 230V 50/60…
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0775-2024·2024-02-07

    FDA recalls EVair 03 compressor models for elevated formaldehyde levels

    FDA recalls EVair 03 air compressors (models 1609000, 1609002) for elevated formaldehyde when used with certain ventilators. The hazard was identified during preliminary testing under worst-case conditions: elevated room temperature (40°C) and minimal gas flow (2 L/min).

    Product
    EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0774-2024·2024-02-07

    EVair Air Compressors Recalled for Elevated Formaldehyde Emissions

    EVair air compressors (models M1230849, M1230847) made by Datex-Ohmeda are being recalled after testing detected elevated formaldehyde emissions. The compressors are used with CARESCAPE R860 and Engström ventilator systems.

    Product
    EVair air compressor, model numbers: M1230849 and M1230847
    Category
    Medical Device
    Distribution
    0 states