Manufacturer

Cardiac Assist, Inc

14 recalls in our database name Cardiac Assist, Inc as the manufacturing or recalling firm.

What the numbers show

Total
14
Critical
0
Severe
2
Most recent
2023-09-06

Of the 14 recalls from Cardiac Assist, Inc we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2 (14%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–14 of 14

  • HighFDA (Devices)·Z-2465-2023·2023-09-06

    LifeSPARC Cardiac Pump Recalled Due to Uncured Epoxy Manufacturing Defect

    Cardiac Assist, Inc. is recalling the LifeSPARC Pump due to uncured epoxy used during manufacturing that may cause early device failure. Two units were distributed in Illinois and Florida.

    Product
    LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray part number: 5800-1000. Note: The 5800-1000 model number is an individually packaged component contained within the Kit numbers: 5830-2916 LS TANDEMLIFE KIT DL29 5840-2417 LS…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0926-2023·2023-01-25

    Medical device controller software update addresses critical failure

    Cardiac Assist is issuing a software update for LifeSPARC Controller units that may experience critical failure causing the system to freeze, crash, or lose display functionality. Affected units are identified by specific serial numbers.

    Product
    SHIP KIT, LIFESPARC, CONTROLLER AND ACCESSORIES - 5900-0000; SHIP KIT, LIFESPARC, CONTROLLER ONLY - 5900-0001. Component of LifeSPARC System, used to pump the blood through an extracorporeal circuit.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1763-2022·2022-10-05

    TANDEM LIFE controller software failures affecting cardiopulmonary bypass equipment nationwide

    FDA recalls TANDEM LIFE LivaNova LIFESPARC Controllers due to critical software failures that cause the device to freeze, crash, or stop displaying data. The controllers are used in cardiopulmonary bypass procedures.

    Product
    TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model 5900-0000; Controller Only Model 5900-0001. For circulatory or cardiopulmonary bypass.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0311-2022·2021-12-08

    Cardiac Assist Procedure Kit Recalled for Incorrect Expiration Date Label

    Cardiac Assist, Inc. is recalling 7 units of the SHIP KIT cardiopulmonary bypass procedure kit due to an incorrect expiration date on the side label. The side label shows May 1, 2021, while the correct expiration date is May 1, 2022.

    Product
    SHIP KIT, PROCEDURE KIT, PROTEKDUO LS - DL31 for temporary Cardiopulmonary bypass. Including the following components: a) 31 Fr ProtekDuo Veno-Venous Cannula Set b) LifeSPARC Pump c) 14Fr to 26 Fr Dilator Set d) TandemHeart Wrap. Kit Number: 5820-3118 Intended use is…
    Category
    Medical Device
    Distribution
    4 states
  • ModerateFDA (Devices)·Z-1486-2021·2021-05-05

    Cardiac Assist Recalls Incorrectly Packaged Protek Solo Venous Dilator Sets

    Cardiac Assist, Inc. is recalling 13 Protek Solo Venous Dilator Sets due to incorrect packaging discovered in finished goods.

    Product
    Protek Solo Venous Dilator Set: dilator, vessel, for percutaneous catheterization. Catalog #: 5840-2417 - LS Tandem Life Kit - V24/A17 Catalog #: 5840-2415 - LS Tandem Life Kit - V24/A15 Catalog #: 5740-2417 - TandemLife Kit - V24/A17 Catalog #: 5100-1422 - ProtekSolo…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1950-2020·2020-05-20

    Cardiac Assist Recalls 15 TandemHeart Pumps Due to Assembly Error

    Cardiac Assist, Inc. is recalling 15 TandemHeart pump kits due to an assembly error that may prevent the device from priming.

    Product
    TandemHeart pump is assembled into kits: TandemHeart Kit - TS62/A17, Product: 5710-6217 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1955-2020·2020-05-20

    Cardiac Assist Recalls 13 TandemLife Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling 13 TandemLife Kits because an assembly error may prevent the pump from priming. The kits provide temporary circulatory support.

    Product
    TandemHeart pump is assembled into kits: TandemLife Kit Product: 5740-0000 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1953-2020·2020-05-20

    Cardiac Assist Recalls Six TandemHeart Pump Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling six TandemHeart pump kits because an assembly error may prevent the device from priming.

    Product
    TandemHeart pump is assembled into kits: ProtekDuo Kit - DL31 RD, Product: 5720-3631 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1948-2020·2020-05-20

    Cardiac Assist Recalls TandemHeart Pump Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling 10 TandemHeart Pump Kits because an assembly error may prevent the device from priming.

    Product
    TandemHeart Pump Kit, Product: 5120-0000 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1954-2020·2020-05-20

    Cardiac Assist Recalls TandemLung Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling two TandemLung Kit units because an assembly error may prevent the pump from priming.

    Product
    TandemHeart pump is assembled into kits: TandemLung Kit - DL31, Product: 5730-3118 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1952-2020·2020-05-20

    Cardiac Assist Recalls TandemHeart ProtekDuo Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling 46 TandemHeart ProtekDuo Kits due to an assembly error that may cause failure to prime.

    Product
    TandemHeart pump is assembled into kits: ProtekDuo Kit - DL3, Product: 5720-3118 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1951-2020·2020-05-20

    Cardiac Assist Recalls 38 TandemHeart ProtekDuo Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling 38 TandemHeart ProtekDuo Kits because an assembly error may prevent the pump from priming.

    Product
    TandemHeart pump is assembled into kits: ProtekDuo Kit - DL29, Product: 5720-2916 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1949-2020·2020-05-20

    Cardiac Assist Recalls TandemHeart Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling five TandemHeart pump kits because an assembly error may prevent the device from priming.

    Product
    TandemHeart pump is assembled into kits: TandemHeart Kit - TS62/A15, Product: 5710-6215 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1106-2020·2020-02-19

    Cardiac Assist Recalls TandemLife ECLS Procedure Kits with Incorrect Cannula

    Cardiac Assist, Inc. is recalling 11 TandemLife ECLS Procedure Kits that contained a 29Fr ProtekDuo cannula instead of the intended 24Fr Protek Solo Venous cannula.

    Product
    TandemLife ECLS Procedure Kit - V24/A17, Catalog Number 5740-2417 - Product Usage: The function of the 24Fr ProtekSolo Venous Cannula is to provide a conduit for blood to travel from the patient s venous system when connected to accessory extracorporeal equipment. The Venous…
    Category
    Medical Device
    Distribution
    Distributed nationwide