The Recall Desk

Manufacturer

Belmont Instrument LLC

6 recalls in our database name Belmont Instrument LLC as the manufacturing or recalling firm.

What the numbers show

Total
6
Critical
0
Severe
0
Most recent
2025-06-18

Of the 6 recalls from Belmont Instrument LLC we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–6 of 6

  • HighFDA (Devices)·Z-1966-2025·2025-06-18

    Allon 2001 Medical Device Heating System Malfunction Risk

    Belmont Instrument's Allon 2001 device may heat water above safe limits if sensor connections fail and temperature alarms are disabled. The patient wrap may exceed its set temperature without user notification.

    Product
    Allon 2001. Part Number: 200-00263
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1678-2025·2025-05-07

    Belmont 3.0 Liter Disposable Set Recalled for Cracked Female Connector

    Belmont Instrument LLC is recalling the 3.0 Liter Disposable Set (Model 903-00032) due to a potential crack in the female connector on the Heat Exchanger/patient line set that can cause fluid leaks during device priming.

    Product
    3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoirs (P/N 903-00018) and 4 Heat Exchanger/Patient Line set (P/N 403-00382). The two components are packaged separately in individual pouches and placed in…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1675-2025·2025-05-07

    Belmont Straight Inflow/Outflow Patient Line Kit recalled for potential connector crack

    Belmont Instrument LLC is recalling the Straight Inflow/Outflow Patient Line Kit (Model 902-00037) due to a potential crack in the female connector on the Heat Exchanger set that may cause fluid leaks during use.

    Product
    Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00037) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1674-2025·2025-05-07

    Hyperthermia Pump Procedure Kit Recalled for Cracked Connector

    Belmont Instrument LLC is recalling 158 Hyperthermia Pump Procedure Kits (Model 902-00045) due to a potential crack in the female connector on the Heat Exchanger set that may cause fluid leakage during use.

    Product
    Hyperthermia Pump Procedure Kit. Model Number: 902-00045. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Hyperthermia Procedure kit (P/N 902-00045) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and HP…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1676-2025·2025-05-07

    Belmont Hyperthermia Pump Line Kit Heat Exchanger Connector Crack Recall

    Belmont Instrument is recalling the Straight Inflow/Bifurcated Outflow Line Kit (Model 902-00038) due to potential cracks in the female connector on the Heat Exchanger set, which may cause fluid leaks during priming.

    Product
    Straight Inflow/Bifurcated Outflow Line Kit. Model Number: 902-00038. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00038) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1099-2025·2025-02-12

    Belmont Medical 3-Spike Disposable Set May Leak During Use

    The Belmont Medical 3-Spike Disposable Set may leak during priming due to a crack in its female quick connector, potentially delaying warmed infusions in hospital settings.

    Product
    Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2. Designed to be used in general operation in hospital or alternate care environments to provide warmed blood and fluids to any patients e10 kg requiring warmed infusion. Part Number:903-00006
    Category
    Medical Device
    Distribution