The Recall Desk

Manufacturer

Aggredyne, Inc.

1 recall in our database name Aggredyne, Inc. as the manufacturing or recalling firm.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2022-09-07

Of the 1 recalls from Aggredyne, Inc. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–1 of 1

  • HighFDA (Devices)·Z-1639-2022·2022-09-07

    AggreGuide A-100 Instrument Software Issue Affecting Test Results

    Aggredyne, Inc. is recalling the AggreGuide A-100 Instrument due to a software design defect that causes the device to run assays at 0 RPM, resulting in no blood mixing and inaccurate test results displayed as Low.

    Product
    AggreGuide A-100 Instrument
    Category
    Medical Device
    Distribution
    0 states