The Recall Desk

Manufacturer

Abbott Diagnostics Technologies AS

2 recalls in our database name Abbott Diagnostics Technologies AS as the manufacturing or recalling firm.

What the numbers show

Total
2
Critical
0
Severe
0
Most recent
2024-09-04

Of the 2 recalls from Abbott Diagnostics Technologies AS we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–2 of 2

  • HighFDA (Devices)·Z-2736-2024·2024-09-04

    Lab analyzer incorrectly configured for complex testing at waived facilities

    Five Afinion 2 lab analyzers were distributed with incorrect configuration, potentially allowing waived laboratories to perform moderately complex assays that could produce erroneous results.

    Product
    Afinion 2 Analyzer, model #1116986, US refurbished, CLIA waived.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2192-2020·2020-06-10

    Abbott Afinion 2 Power Supply Recall for Electrostatic Discharge Risk

    Abbott is recalling Afinion 2 and Alere Afinion AS100 power supplies due to a risk of electrostatic discharge and sparking that can render the unit non-functional.

    Product
    Power supply provided with Afinion 2, Catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, 1116679, 1116680, 1116681, 1116682, 1116684, 1116770, 1116771, 1116772, 1116777, 1116778, 1116970, 1116971, 1116985, 1116986, IVD. The firm name on the label is A
    Category
    Medical Device
    Distribution
    1 state