The Recall Desk

Hazard

Wrong Size Administration recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2016-03-09

Of the 1 wrong size administration recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all wrong size administration recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • ModerateFDA (Devices)·Z-1034-2016·2016-03-09

    FDA Recalls Mispackaged Pediatric Feeding Tubes from Covidien

    The FDA is recalling 1,330 units of pediatric feeding tubes due to packaging errors where different tube sizes were placed in incorrect packages.

    Product
    Covidien Kangaroo Polyurethane Feeding Tube Radiopaque Line, Safe Enteral Connections 8 Fr/Ch (2.7 mm) x 20" (51 cm); Catalog Number: 461701. Intended for pediatric feeding.
    Category
    Medical Device
    Distribution
    1 state