The Recall Desk

Hazard

Weight Capacity Mismatch recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
1
Most recent
2020-11-10

Of the 1 weight capacity mismatch recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1 (100%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all weight capacity mismatch recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • SevereNHTSA·20V695000·2020-11-10

    2021 Winnebago Micro Minnie Trailers Undersized Axles Recall

    Winnebago is recalling certain 2021 Micro Minnie trailers equipped with undersized axles that may break under the vehicle's weight, potentially causing serious injury or death.

    Product
    WINNEBAGO — WINNEBAGO MICRO MINNIE
    Category
    Vehicle
    Distribution
    Distributed nationwide