The Recall Desk

Hazard

Toxic Liquid recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
1
Most recent
2001-06-19

Of the 1 toxic liquid recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1 (100%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all toxic liquid recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • SevereCPSC·01166·2001-06-19

    CPSC Recalls Remote-Controlled Toy Cars and Galileo Weather Thermometers

    CPSC and firms recall remote-controlled toy cars and Galileo thermometers due to fire and burn hazards. Consumers should stop using these items immediately.

    Product
    Remote Controlled Toy Race Cars and Galileo Weather Thermometers
    Category
    Consumer Product
    Distribution
    Distributed nationwide