The Recall Desk

Hazard

Tire Information Mismatch recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2020-08-27

Of the 1 tire information mismatch recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all tire information mismatch recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • ModerateNHTSA·20V514000·2020-08-27

    2021 BMW X7 Recalled for Incorrect Tire Information Labels

    BMW is recalling 2021 X7 models for tire information labels with incorrect specifications. The labels may not match the actually installed tires, potentially causing drivers to select wrong tires.

    Product
    BMW — BMW X7
    Category
    Vehicle
    Distribution
    Distributed nationwide