The Recall Desk

Hazard

Tip Detachment recalls

20 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
20
Critical
7
Severe
6
Most recent
2025-10-22

Of the 20 tip detachment recalls we have scored against our rubric, 7 (35%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 6 (30%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all tip detachment recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–20 of 20

  • CriticalFDA (Devices)·Z-0112-2026·2025-10-22

    Curve F-J ablation catheters recalled after tip detachment during unpacking

    Abbott is recalling 11,810 TactiFlex Sensor Enabled Ablation Catheters in the F-J curve after becoming aware of tip detachment events during removal from the packaging.

    Product
    TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, FJ, (BiD Curve F-J), REF A-TFSE-FJ; cardiac catheter
    Category
    Medical Device
    Distribution
    0 states
  • CriticalFDA (Devices)·Z-0109-2026·2025-10-22

    UniD curve ablation catheters recalled after tip detachment from packaging

    Abbott is recalling 407 TactiFlex Sensor Enabled Ablation Catheters in the UniD curve D configuration after becoming aware of tip detachment events during removal from the packaging.

    Product
    TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, D, (UniD Curve D), REF A-TFSE-D; cardiac catheter
    Category
    Medical Device
    Distribution
    0 states
  • CriticalFDA (Devices)·Z-0113-2026·2025-10-22

    UniD curve F ablation catheters recalled after tip detachment from packaging

    Abbott is recalling 1,960 TactiFlex Sensor Enabled Ablation Catheters in the UniD curve F configuration after becoming aware of tip detachment events during removal from the packaging.

    Product
    TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, F, (UniD Curve F), REF A-TFSE-F; cardiac catheter
    Category
    Medical Device
    Distribution
    0 states
  • CriticalFDA (Devices)·Z-0110-2026·2025-10-22

    Curve D-D ablation catheters recalled after tip detachment from packaging

    Abbott is recalling 3,404 TactiFlex Sensor Enabled Ablation Catheters in the D-D curve after becoming aware of tip detachment events during removal from the packaging.

    Product
    TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DD, (BiD Curve D-D), REF A-TFSE-DD; cardiac catheter
    Category
    Medical Device
    Distribution
    0 states
  • CriticalFDA (Devices)·Z-0111-2026·2025-10-22

    Curve D-F ablation catheters recalled after tip detachment from packaging

    Abbott is recalling 85,803 TactiFlex Sensor Enabled Ablation Catheters in the D-F curve after becoming aware of tip detachment events during removal from the packaging.

    Product
    TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DF, (BiD Curve D-F), REF A-TFSE-DF; cardiac catheter
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0129-2026·2025-10-15

    Shorter 8 French feeding tubes recalled because distal tips may detach

    Applied Medical Technology is recalling 840 NutraGlide 5 F - 8 F x 55 nasal feeding tubes whose distal tips may detach at lower than expected forces.

    Product
    NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 8 F x 55, REF E NRT-06055-I The NutraGlide Nasal Feeding Tube is a feeding tube placed through the nose and ending in the stomach or intestine used for the administration of nutrition, fluids, and medications in…
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0128-2026·2025-10-15

    Nasal feeding tubes recalled because distal tips may detach too easily

    Applied Medical Technology is recalling 390 NutraGlide nasal feeding tubes whose distal tips may detach at lower than expected forces.

    Product
    NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 90, REF E NRT-05090-I The NutraGlide Nasal Feeding Tube is a feeding tube placed through the nose and ending in the stomach or intestine used for the administration of nutrition, fluids, and medications in…
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0127-2026·2025-10-15

    Shorter 6 French feeding tubes recalled because distal tips may detach

    Applied Medical Technology is recalling 1,780 NutraGlide 5 F - 6 F x 55 nasal feeding tubes whose distal tips may detach at lower than expected forces.

    Product
    NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 55, REF E NRT-05055-I The NutraGlide Nasal Feeding Tube is a feeding tube placed through the nose and ending in the stomach or intestine used for the administration of nutrition, fluids, and medications in…
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0130-2026·2025-10-15

    Larger nasal feeding tubes recalled because distal tips may detach

    Applied Medical Technology is recalling 200 NutraGlide 5 F - 8 F nasal feeding tubes whose distal tips may detach at lower than expected forces.

    Product
    NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 8 F x 90, REF E NRT-06090-I The NutraGlide Nasal Feeding Tube is a feeding tube placed through the nose and ending in the stomach or intestine used for the administration of nutrition, fluids, and medications in…
    Category
    Medical Device
    Distribution
    2 states
  • SevereFDA (Devices)·Z-1512-2025·2025-04-16

    Hippo 072 Aspiration System recalled for distal tip detachment risk

    Qapel Medical Inc. is recalling the Hippo 072 Aspiration System and related components because the aspiration catheter distal tip may detach, potentially causing vessel rupture or vasospasm.

    Product
    Hippo 072 Aspiration System including Cheetah Delivery Tool and Aspiration Tubing, REF: APT6072-132. Hippo 072 Aspiration System including Cheetah Delivery Tool, REF: AP6072-132. 072 Aspiration Tubing, REF: APT-95
    Category
    Medical Device
    Distribution
    29 states
  • SevereFDA (Devices)·Z-0694-2022·2022-03-16

    Medtronic TurboHawk Plus Directional Atherectomy System Recalled Due to Design Defect

    ev3 Inc. is recalling the Medtronic TurboHawk Plus Directional Atherectomy System due to design similarities to a device with reported tip detachment and embolization incidents. The FDA has classified this as a Class I recall.

    Product
    Medtronic TurboHawk Plus Directional Atherectomy System 6F, Multi-vessel, Catalog number THP-M. For use in peripheral vasculature.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0695-2022·2022-03-16

    Medtronic TurboHawk Plus atherectomy system recalled for potential tip detachment

    The Medtronic TurboHawk Plus Directional Atherectomy System (catalog THP-S) is being recalled due to design similarities to another device that experienced tip detachment and embolization. The recall affects 245 devices distributed nationwide.

    Product
    Medtronic TurboHawk Plus Directional Atherectomy System 6F, Small Vessel, Catalog number THP-S. For use in peripheral vasculature.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • CriticalFDA (Devices)·Z-0450-2022·2022-01-26

    Medtronic HawkOne Atherectomy System Recalled for Tip Detachment and Embolization Risk

    Medtronic is recalling the HawkOne Directional Atherectomy System globally due to reports of tip damage that can detach and cause embolization. The device is used to remove arterial plaque in peripheral blood vessels.

    Product
    Medtronic HawkOne Directional Atherectomy System, Catalog numbers H1-M and H1-M-INT., Multi-vessel, 6F (2.2 mm) Recommended sheath, 0.014" (0.36 mm) Compatible guidewire. The HawkOne directional atherectomy system is intended for use in atherectomy of the peripheral…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0451-2022·2022-01-26

    Medtronic HawkOne Directional Atherectomy Device Recalled for Tip Damage

    Medtronic is recalling 28,094 units of its HawkOne directional atherectomy system due to reports of catheter tip damage and detachment that can cause blood clots in blood vessels.

    Product
    Medtronic HawkOne Directional Atherectomy System, Catalog numbers H1-S and H1-S-INT, Small Vessel, 6F (2.2 mm) Recommended sheath, 0.014" (0.36 mm) Compatible guidewire. The HawkOne directional atherectomy system is intended for use in atherectomy of the peripheral vasculature.…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2330-2021·2021-09-01

    AQUABEAM Handpiece scope tube tip may detach during procedures

    PROCEPT BIOROBOTICS is recalling AQUABEAM Handpieces (model HP2000) because the scope tube tip may detach during use, potentially requiring procedural delay or additional intervention to remove the tip from the bladder or prostate.

    Product
    AQUABEAM Handpiece, REF: HP2000, a sterile single-use component of the AQUABEAM Robotic System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0741-2018·2018-03-07

    Cosmedent Etching Gel Recalled Due to Syringe Tip Detachment Risk

    Cosmedent, Inc. is recalling 84 units of Etching Gel because the syringe tips may detach during use.

    Product
    Cosmedent's Etching Gel, 3g, 37% Phosphoric Acid, Rx only. Packaged in 3g Luer Lock syringes. Used for etching enamel and dentin.
    Category
    Medical Device
    Distribution
    5 states
  • SevereFDA (Devices)·Z-1898-2017·2017-05-24

    ReFlow Medical Recalls Wingman 35 Crossing Catheters Due to Tip Detachment Risk

    ReFlow Medical is recalling Wingman 35 Crossing Catheters because they may be prone to tip detachment. The recall affects 2,794 units distributed in the US and internationally.

    Product
    Wingman 35 Crossing Catheter, 65cm, Model Numbers: WGM35065US WGM35065CE WGM35090US WGM35090CE WGM35135US WGM35135CE
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1747-2014·2014-06-18

    Neo Metrics Recalls NovaGold Guidewires Due to Tip Uncoiling Risk

    Neo Metrics, Inc. is recalling NovaGold High Performance Guidewires because the tip may uncoil or detach more frequently than anticipated. No illnesses or injuries have been reported.

    Product
    Boston Scientific, NovaGold High Performance Guidewire, Rx only, Sterilized using ethylene oxide, UPN M00552010 and M00552000, Manufactured by NeoMetrics Inc, 2605 Fernbrook Lane Suite J, Plymouth, MN 55447. The NovaGold Guidewire is intended for use in selective cannulation…
    Category
    Medical Device
    Distribution
    0 states