The Recall Desk

Hazard

Steam Escape recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
1
Most recent
2026-04-09

Of the 1 steam escape recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1 (100%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all steam escape recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • SevereCPSC·26385·2026-04-09

    BISSELL steam cleaner attachments can detach and expel hot water

    BISSELL Steam Shot OmniReach and Omni handheld steam cleaners are recalled after 206 reports of hot water or steam escaping from the attachments, including 161 reports of burn injuries.

    Product
    Steam Shot OmniReach and Steam Shot Omni Steam Cleaners with attachments
    Category
    Consumer Product
    Distribution
    Distributed nationwide