The Recall Desk

Hazard

Speed Non Compliance recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2021-09-29

Of the 1 speed non compliance recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all speed non compliance recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighCPSC·21204·2021-09-29

    Luyuan Youth ATVs Recalled for Crash Hazard and Safety Standard Violation

    Luyuan Inc. is recalling about 6,800 youth all-terrain vehicles sold from August 2018 through August 2020 because they fail to comply with federal ATV safety standards, including maximum speed limitations required for children under 10.

    Product
    Luyuan Youth All-Terrain Vehicles (ATVs)
    Category
    Vehicle
    Distribution
    Distributed nationwide