The Recall Desk

Hazard

Shelf Life recalls

74 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
74
Critical
0
Severe
0
Most recent
2026-04-15

Of the 74 shelf life recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all shelf life recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–25 of 74

  • HighFDA (Devices)·Z-1782-2026·2026-04-15

    Cook Medical extubation set labeled with an overstated expiration date

    Cook Medical is recalling Staged Extubation Sets because affected lots were labeled with expiration dates that exceed the product's true shelf life. 95 units are affected.

    Product
    COOK MEDICAL Cook Staged Extubation Set: Reference Part Number C-CAE-14.0-83-EXTU, Order Number G24073
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1792-2026·2026-04-15

    Cook Medical pneumothorax tray labeled with an overstated expiration date

    Cook Medical is recalling Wayne Pneumothorax Trays because affected lots were labeled with expiration dates that exceed the product's true shelf life. 318 units are affected.

    Product
    COOK MEDICAL Wayne Pneumothorax Tray: Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1784-2026·2026-04-15

    Cook Medical bile duct exploration sets labeled with overstated expiration dates

    Cook Medical is recalling the Strange Bile Duct Stone Exploration Set because affected lots were labeled with expiration dates that exceed the product's true shelf life. 49 units are affected.

    Product
    COOK MEDICAL Cook¿ Strange Bile Duct Stone Exploration Set: Reference Part Number C-CSRC-7.5D-40-STRANGE, Order Number G09108
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1777-2026·2026-04-15

    Cook Medical CTO-18 wire guides labeled with overstated expiration dates

    Cook Medical is recalling Approach CTO-18 Micro Wire Guides because affected lots were labeled with expiration dates that exceed the product's true shelf life. 75 units are affected.

    Product
    COOK MEDICAL Approach CTO-18 Micro Wire Guide, Reference Part Number CMW-14-190-18G, Order Number G50788; Reference Part Number CMW-14-300-18G, Order Number G50792
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1788-2026·2026-04-15

    Cook Medical central venous catheter sets labeled with overstated expiration dates

    Cook Medical is recalling Spectrum Central Venous Catheter Sets because affected lots were labeled with expiration dates that exceed the product's true shelf life. 5,080 units are affected.

    Product
    COOK MEDICAL Spectrum¿ Central Venous Catheter Set: Reference Part Number C-UDLM-401J-ABRM-HC, Order Number G43983; Reference Part Number C-UDLM-401J-ABRM-HC-CAH, Order Number G24754; Reference Part Number C-UDLM-401J-LSC-ABRM-HC, Order Number G43986; Reference Part…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1780-2026·2026-04-15

    Blue Rhino G2-Multi tracheostomy introducer trays carry overstated expiration dates

    Cook Medical is recalling Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Trays because affected lots were labeled with expiration dates that exceed the true shelf life. 441 units are affected.

    Product
    COOK MEDICAL Blue Rhino¿ G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-HC-G-NA, Order Number G57706; Reference Part Number C-PTISY-100-UNS-HC-G-NA, Order Number…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1783-2026·2026-04-15

    Cook Medical Spectrum central venous catheter trays carry overstated expiration dates

    Cook Medical is recalling custom Spectrum Central Venous Catheter Trays because affected lots were labeled with expiration dates that exceed the true shelf life. 1,872 units are affected.

    Product
    COOK MEDICAL Cook¿ Spectrum¿ Central Venous Catheter Tray: Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0003, Order Number G56424; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0018, Order Number G21053; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0042, Order Number…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1789-2026·2026-04-15

    Cook Medical Spectrum central venous trays carry overstated expiration dates

    Cook Medical is recalling Spectrum Central Venous Trays because affected lots were labeled with expiration dates that exceed the product's true shelf life. 2,348 units are affected.

    Product
    COOK MEDICAL Spectrum¿ Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order Number G44800; Reference Part Number C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, Order Number G44809; Reference Part Number C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, Order Number…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1790-2026·2026-04-15

    Cook Medical chest tube tray labeled with an overstated expiration date

    Cook Medical is recalling Thal-Quick Chest Tube Trays because affected lots were labeled with expiration dates that exceed the product's true shelf life. Two units are affected.

    Product
    COOK MEDICAL Thal-Quick Chest Tube Tray: Reference Part Number C-TQTSY-3200, Order Number G07242
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1785-2026·2026-04-15

    Cricothyrotomy catheter sets labeled with overstated expiration dates

    Cook Incorporated is recalling Melker Universal Cuffed Emergency Cricothyrotomy Catheter Sets because affected lots were labeled with expiration dates that exceed the true shelf life.

    Product
    Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part Number C-TCCSB-500-UNI-01, Order Number G47665.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1779-2026·2026-04-15

    Blue Rhino G2-Multi tracheostomy introducer sets carry overstated expiration dates

    Cook Medical is recalling Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Sets because affected lots were labeled with expiration dates that exceed the true shelf life. 916 units are affected.

    Product
    COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-HC-G-EU, Order Number G57685; Reference Part Number C-PTIS-100-HC-G-NA, Order Number G57682; Reference Part Number C-PTIS-100-UNL-HC-G-EU, Order Number G57687…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1791-2026·2026-04-15

    Cook Medical Wayne pneumothorax sets carry overstated expiration dates

    Cook Medical is recalling Wayne Pneumothorax Sets because affected lots were labeled with expiration dates that exceed the product's true shelf life. 104 units are affected.

    Product
    COOK MEDICAL Wayne Pneumothorax Set: Reference Part Number C-UTPT-1020-WAYNE-IMH, Order Number G56532; Reference Part Number C-UTPT-1400-WAYNE-112497-IMH, Order Number G56535.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1778-2026·2026-04-15

    Cook Medical arterial pressure monitoring trays carry overstated expiration dates

    Cook Medical is recalling Arterial Pressure Monitoring Trays because affected lots were labeled with expiration dates that exceed the product's true shelf life. 20 units are affected.

    Product
    COOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1786-2026·2026-04-15

    Cook Medical multipurpose drainage catheters carry overstated expiration dates

    Cook Medical is recalling a Multipurpose Drainage Catheter because affected lots were labeled with expiration dates that exceed the product's true shelf life. 20 units are affected.

    Product
    COOK MEDICAL Multipurpose Drainage Catheter: Reference Part Number ULT8.5-38-25-P-CLMS-8-RH-FS, Order Number G29775
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1781-2026·2026-04-15

    Cook Medical tracheostomy introducer sets labeled with overstated expiration dates

    Cook Medical is recalling Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Sets because affected lots were labeled with expiration dates that exceed the true shelf life. 484 units are affected.

    Product
    COOK MEDICAL Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-FLEX-HC, Order Number G32092; Reference Part Number C-PTIS-100-HC, Order Number G12115.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1787-2026·2026-04-15

    Cook Medical nitinol stone extractors labeled with overstated expiration dates

    Cook Medical is recalling NCompass Nitinol Stone Extractors because affected lots were labeled with expiration dates that exceed the product's true shelf life. 70 units are affected.

    Product
    COOK MEDICAL NCompass¿ Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250; Reference Part Number C-NTSE-2.4-115-NCT4, Order Number G36251.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1776-2026·2026-04-15

    Cook Medical micro wire guides labeled with overstated expiration dates

    Cook Medical is recalling Approach CTO-12 Micro Wire Guides because affected lots were labeled with expiration dates that exceed the product's true shelf life. 49 units are affected.

    Product
    COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-1492-2026·2026-03-11

    Vascular grafts recalled after expiry dates were calculated from packaging

    Vascutek is recalling Gelsoft Plus and Gelweave vascular prostheses because they were labeled with an additional month of shelf life, expiry having been calculated from packaging rather than gel impregnation.

    Product
    Gelsoft Plus, REF: 635024PE, 631514PE, 631222PE, 633008PE, 633010PE, 633007PE, 636010PE, 632522PE, 633012PE, 635018PE, 631516PE, 636016PE, 631220PE, 631608PE, 631206PE, 631407P50E, 631508PE, 636008PE, 636006PE, 632518PE, 634008PE, 634006PE, 631218PE, 632211PE, 634007PE…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0897-2026·2025-12-17

    Chemotherapy gowns recalled over an expiration date printed two years late

    Cardinal Health is recalling 251,165 ChemoPlus gowns and sleeves because the packaging shows a five-year shelf life when the product's actual shelf life is three years.

    Product
    ChemoPlus gowns: Product Code Product Description CT5500T ChemoPlus Full Coverage Sleeve, Universal, Medium Blue, Non-sterile; CT5502T ChemoPlus Full Coverage Gown, Closed Back, Regular/Medium, Medium Blue, Non-sterile; CT5503T ChemoPlus Full Coverage Gown, Closed Back…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2581-2025·2025-09-24

    Sixteen millimetre distraction pins recalled over unvalidated shelf life labeling

    Frontier Devices 16 mm Distraction Pins in 10 single packs are recalled because the labeling includes a shelf life that has not been validated. 490 units are affected.

    Product
    Frontier Devices, REF: 301.916S1, 16 mm Distraction Pin, 10 single packs , Rx Only, Sterile
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2578-2025·2025-09-24

    Distraction pins in five packs recalled over unvalidated shelf life

    Frontier Devices 14 mm Distraction Pins in 5 single packs are recalled because the labeling includes a shelf life that has not been validated. 2,700 units are affected.

    Product
    Frontier Devices, REF: 301.914S1, 14 mm Distraction Pin, 5 single packs , Rx Only, Sterile
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2579-2025·2025-09-24

    Fourteen millimetre distraction pins recalled over unvalidated shelf life labeling

    Frontier Devices 14 mm Distraction Pins in 10 single packs are recalled because the labeling includes a shelf life that has not been validated. 10,960 units are affected.

    Product
    Frontier Devices, REF: 301.914S1, 14 mm Distraction Pin, 10 single packs , Rx Only, Sterile
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2577-2025·2025-09-24

    Twelve millimetre distraction pins recalled over unvalidated shelf life labeling

    Frontier Devices 12 mm Distraction Pins in 10 double packs are recalled because the labeling includes a shelf life that has not been validated. 160 units are affected.

    Product
    Frontier Devices, REF: 301.912S2, 12 mm Distraction Pin, 10 double packs , Rx Only, Sterile
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2580-2025·2025-09-24

    Distraction pins in double packs recalled over unvalidated shelf life labeling

    Frontier Devices 14 mm Distraction Pins in 10 double packs are recalled because the labeling includes a shelf life that has not been validated.

    Product
    Frontier Devices, REF: 301.914S2, 14 mm Distraction Pin, 10 double packs , Rx Only, Sterile
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2576-2025·2025-09-24

    Distraction pins in single packs recalled over shelf life that was not validated

    Frontier Devices 12 mm Distraction Pins in 10 single packs are recalled because the labeling includes a shelf life that has not been validated. 3,790 units are affected.

    Product
    Frontier Devices, REF: 301.912S1, 12 mm Distraction Pin, 10 single packs , Rx Only, Sterile
    Category
    Medical Device
    Distribution
    Distributed nationwide