Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all sample destruction recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighFDA (Devices)·Z-2980-2024·2024-09-11

    Beckman Power Express Centrifuge Sample Holder May Detach During Operation

    An adhesive joint failure in the PE Centrifuge Temperature Controlled component can cause the metal part to detach inside the centrifuge, risking sample destruction and potential exposure of laboratory technicians to biohazardous material.

    Product
    PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.
    Category
    Medical Device
    Distribution
    Distributed nationwide