The Recall Desk

Hazard

Safety Interlock Bypass recalls

3 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
3
Critical
0
Severe
0
Most recent
2023-08-16

Of the 3 safety interlock bypass recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all safety interlock bypass recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–3 of 3

  • HighFDA (Devices)·Z-2397-2023·2023-08-16

    Philips MR 7700 Magnetic Resonance System Smoke Detector Interlock Bypass

    Philips MR 7700 MRI systems can have their smoke detector interlock safety mechanism bypassed when power cycled after smoke detection, allowing continued operation despite potential fire hazards.

    Product
    MR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2395-2023·2023-08-16

    MR Imaging System Smoke Detector Interlock Vulnerable to Bypass via Power Cycling

    Philips Ingenia Elition X MR systems can have their smoke detection safety interlock bypassed by power cycling the equipment. This vulnerability could allow unsafe operation during hazardous conditions.

    Product
    Ingenia Elition X Magnetic Resonance (MR) system, Product Numbers 781358, 782107, and 782118 (SmartPath to Ingenia Elition X)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2396-2023·2023-08-16

    Ingenia Elition S MR System smoke detection interlock vulnerable to operator bypass

    Philips Ingenia Elition S MR systems can be circumvented by power cycling to bypass the smoke detection safety interlock that locks after smoke is detected. This could allow continued operation despite fire hazards.

    Product
    Ingenia Elition S Magnetic Resonance (MR) system, Product Numbers 781357 and 782106
    Category
    Medical Device
    Distribution
    Distributed nationwide