DxI 600 with Spot B analyzer motor current resets below requirement
DxI 600 Access Immunoassay Analyzers with Spot B are being recalled because software overwrites the factory PnP X-motor current setting, causing increased motion errors. 177 systems are affected.
- Product
- DxI 600 Access Immunoassay Analyzer W/Spot B, Part Number A71460
- Category
- Medical Device
- Distribution
- Distributed nationwide