The Recall Desk

Hazard

Reprocessing Incompatibility recalls

5 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
5
Critical
0
Severe
0
Most recent
2026-04-01

Of the 5 reprocessing incompatibility recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all reprocessing incompatibility recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–5 of 5

  • HighFDA (Devices)·Z-1612-2026·2026-04-01

    Olympus OER-Elite reprocessor no longer compatible with two valve models

    Olympus is recalling the OER-Elite Endoscope Reprocessor because the MAJ-1443 and MAJ-1444 air/water valves are no longer reprocessing compatible with the OER-Pro and OER-Elite reprocessors.

    Product
    Olympus Endoscope Reprocessor OER-Elite. Model Number: N5789340.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1613-2026·2026-04-01

    Olympus OER-Pro reprocessor no longer compatible with two valve models

    Olympus is recalling the OER-Pro Endoscope Reprocessor because the MAJ-1443 and MAJ-1444 air/water valves are no longer reprocessing compatible with the OER-Pro and OER-Elite reprocessors.

    Product
    Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1610-2026·2026-04-01

    Olympus endoscope suction valve no longer compatible with named reprocessors

    Olympus is recalling the MAJ-1443 Endoscope Suction Valve because the MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

    Product
    Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1611-2026·2026-04-01

    Olympus endoscope air water valves no longer compatible with named reprocessors

    Olympus is recalling the MAJ-1444 Endoscope Air/Water Valve because the MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

    Product
    Olympus Endoscope Air/Water Valve - MAJ-1444. Model Number: MAJ-1444.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1034-2022·2022-05-11

    Ultrasonic Gastrovideoscope Recalled Due to Reprocessing Incompatibility and Infection Control Risk

    Olympus recalls the GF-UC140P-AL5 ultrasonic gastrovideoscope nationwide because it is no longer compatible with the OER-Pro reprocessing system, creating an infection control risk to patients.

    Product
    Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging.
    Category
    Medical Device
    Distribution
    Distributed nationwide