Hematology Analyzer Sight OLO Displays Incorrect Pediatric Reference Ranges
The Sight OLO automated hematology analyzer displays mixed-up default reference ranges for children ages 2-11 in software versions 2.63, 2.63.1, and 2.63.1.1, which could cause test results to appear incorrectly normal or abnormal.
- Product
- Sight OLO, Automated Hematology Analyzer, Catalog No. OLO-U1
- Category
- Medical Device
- Distribution
- Distributed nationwide