The Recall Desk

Hazard

Radiofrequency Interference recalls

3 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
3
Critical
0
Severe
0
Most recent
2023-03-15

Of the 3 radiofrequency interference recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all radiofrequency interference recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–3 of 3

  • HighFDA (Devices)·Z-1219-2023·2023-03-15

    CardioMEMS Heart Monitoring System Emissions May Interfere With Pacemakers

    FDA recalls CardioMEMS HF System patient monitor for radiofrequency signal emissions that can interfere with pacemakers, defibrillators, and other implanted medical devices. The emissions exceed radiofrequency limits.

    Product
    CardioMEMS HF System Patient Electronics System (PES), Model number CM1010, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1221-2023·2023-03-15

    CardioMEMS Hospital Electronics System RF Emissions May Interfere With Implanted Devices

    St. Jude Medical is recalling the CardioMEMS HF System (Model CM3000) due to excessive radiofrequency emissions that may interfere with pacemakers, defibrillators, and other implanted devices during use. The recall affects 1,905 units distributed worldwide.

    Product
    CardioMEMS HF System Hospital Electronics System (HES), Model number CM3000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1218-2023·2023-03-15

    CardioMEMS Heart Failure Monitor Recalled for Radiofrequency Emissions

    St. Jude Medical is recalling certain CardioMEMS heart failure monitoring devices due to excessive radiofrequency emissions that could interfere with pacemakers, defibrillators, and other implanted medical devices when in close proximity.

    Product
    CardioMEMS HF System Patient Electronics System (PES), Model number CM1000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which…
    Category
    Medical Device
    Distribution
    Distributed nationwide