The Recall Desk

Hazard

Pressure Relief Valve Defect recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2021-09-01

Of the 1 pressure relief valve defect recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all pressure relief valve defect recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighFDA (Devices)·Z-2347-2021·2021-09-01

    MRI System Pressure Relief Valve Installation Error May Release Helium Gas

    Philips MRI systems may have incorrect pressure relief valves installed, risking helium gas release and injury. Seven units contain 42 psi valves instead of required 3.75 psi valves.

    Product
    Ingenia Elition S-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model…
    Category
    Medical Device
    Distribution
    Distributed nationwide