Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all pressure relief defect recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighFDA (Devices)·Z-1401-2024·2024-04-10

    Cardiovascular Procedure Kits Recalled for Pressure Relief Valve Defect

    Terumo is recalling 20,533 cardiovascular procedure kits due to a pressure relief valve that opens at incorrect pressure (200-300 mmHg instead of 400 mmHg). The defect could cause myocardial tissue damage during cardiac surgery.

    Product
    Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974 65246 66312 66342 66388 66650 66852 70472 74307 74365 74366 74430 74475 74477 74502 74773 74873 75065 75157 75195 75409 75410 75473 75533 75534
    Category
    Medical Device
    Distribution
    Distributed nationwide