The Recall Desk

Hazard

Ph Standard Failure recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2015-04-29

Of the 1 ph standard failure recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all ph standard failure recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • ModerateFDA (Food)·F-2021-2015·2015-04-29

    Summit Import Recalls Preserved Duck Eggs for pH and Salt Standard Failures

    Summit Import Corporation is recalling Yang Sheng Preserved Duck Eggs because the product failed to meet required pH and salt standards.

    Product
    Yang Sheng Preserved Egg (Long-Lansing Duck Egg), each box contains 6 preserved eggs, each box weighs about 12 ounce (340g), Manufacturer: Ya Nong Co., Ltd, No. 181-1, Singdong Rd., Jhongli City, Taoyuan County, 320, Taiwan R.O.C., Product of Taiwan
    Category
    Food
    Distribution
    4 states