Radiotherapy systems recalled over software allowing more than two daily fractions
RefleXion X1 systems are recalled because an incorrect software configuration potentially allows more than two treatment fractions within a 12-hour period, which could lead to toxicity. 11 units are affected.
- Product
- RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007…
- Category
- Medical Device
- Distribution
- Distributed nationwide