Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all mri software defect recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighFDA (Devices)·Z-2459-2024·2024-08-14

    SIGNA Architect MRI System Software Can Exceed Implant Heating Limits

    GE Medical Systems is recalling SIGNA Architect MRI systems in which software may allow B1+ radiofrequency exposure to exceed user-prescribed limits during certain T2 FLAIR imaging sequences, risking MR-conditional implant overheating.

    Product
    SIGNA Architect, with affected software versions: DV26.0 to DV26.3, DV27.0 to DV27.3, DV28.0 to DV28.5, DV29.0, DV29.1, DV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states