The Recall Desk

Hazard

Mis Labeling recalls

9,611 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
9,611
Critical
161
Severe
589
Most recent
2026-09-24

Of the 9,611 mis labeling recalls we have scored against our rubric, 161 (2%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 589 (6%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all mis labeling recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

6751–6775 of 9611

  • ModerateFDA (Devices)·Z-0675-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BLUNT FORCEPS, BLUNT, SERRATED, SPOON, P/N 0250080313. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0764-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, SCHERTL GRASPER, FENESTRATED, 20CM, P/N 0250282006 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequen
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0654-2015·2014-12-24

    Stryker Endoscopy Recalls Tissue Grasper Forceps Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of tissue grasper forceps because the instructions for use do not support two specific sterilization methods.

    Product
    PKG, GRASPER FORCEPS, TISSUE, P/N 0250080284. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which a
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0682-2015·2014-12-24

    Stryker Endoscopy Recalls 33cm Dorsey Grasper Forceps Due to Sterilization Labeling

    Stryker Endoscopy is recalling 180,573 units of 33cm Dorsey Grasper Forceps because the instructions for use do not support specific sterilization methods.

    Product
    33CM DORSEY GRASPER FORCEPS, P/N 0250080319. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which all
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0785-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, MARYLAND, FORCEPS CURVED, 29CM, P/N 0250282058 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0664-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Obturator Packages Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of 11mm Obturator packages because the instructions for use do not support specific sterilization methods.

    Product
    PKG, OBTURATOR, 11MM, FOR HASSON CANNULA, P/N 0250080158. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) p
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0742-2015·2014-12-24

    Stryker Endoscopy Recalls Park Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling specific Park Forceps instruments because the instructions for use do not support certain sterilization methods.

    Product
    PKG, 5MM INSERT, 45CM, PARK FORCEPS, P/N 0250080745 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports wh
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0602-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, PORTIO ACORN SMALL ONLY, P/N 0250080139; Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which al
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0778-2015·2014-12-24

    Stryker Endoscopy Recalls Fundus Grasper Instruments Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 Fundus Grasper instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, FUNDUS GRASPER, 29CM, P/N 0250282050 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0787-2015·2014-12-24

    Stryker Endoscopy Recalls Biopsy Punch Instruments Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 biopsy punch instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, BIOPSY PUNCH, 29CM, P/N 0250282060 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports wh
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0730-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Issues

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, ASPIRATION NEEDLE, 33CM, P/N 0250080702 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which all
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0620-2015·2014-12-24

    Stryker Maryland Forceps Recalled Due to Sterilization Parameter Errors

    Stryker Endoscopy is recalling Maryland Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MARYLAND FORCEPS, CURVED, LEFT, P/N 0250080042. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0765-2015·2014-12-24

    Stryker Endoscopy Recalls 180,573 Cobra Grasper Instruments Due to Sterilization Labeling

    Stryker Endoscopy is recalling 180,573 Cobra Grasper instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, COBRA GRASPER, 20CM, P/N 0250282007 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports w
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0781-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, SCHERTL GRASPER, FENESTRATED, 29CM, P/N 0250282053 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequen
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0779-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, FUNDUS GRASPER, 29CPKG, 3MM INSERT, ALLIGATOR GRASPER, 29CM, P/N 0250282051 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instr
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0794-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Issues

    Stryker Endoscopy is recalling laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG., 3MM INSERT, BABCOCK FORCEPS, 29CM, P/N 0250282103 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) port
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0605-2015·2014-12-24

    Stryker Laparoscopic Cannula Recall Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic cannulas because the instructions for use do not support specific sterilization methods.

    Product
    PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O TROCAR, P/N 0250080171. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio f
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0592-2015·2014-12-24

    Stryker Endoscopy Recalls Bipolar Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of bipolar forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BIPOLAR FORCEPS, MICRO TIP, 45CM, P/N 0250080115 nationwide in US and worldwide: Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0708-2015·2014-12-24

    Stryker Endoscopy Recalls 5mm Insert Biopsy Forceps Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of 5mm Insert Biopsy Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT BIOPSY FORCEPS, SPOON W/PIN, 45CM, P/N 0250080581 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0645-2015·2014-12-24

    Stryker Mini Metzenbaum Scissors Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Mini Metzenbaum Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MINI METZENBAUM SCISSORS, CURVED LONG, P/N 0250080266. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0798-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Instruments Due to Sterilization Labeling Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, CANNULA 8MM AUTOMATIC VALVE STOP COCK, W/O TROCAR, P/N 0250080172 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio freq
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0751-2015·2014-12-24

    Stryker Laparoscopic Dissector Recall Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 laparoscopic dissectors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM X 45CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW, P/N 0250080761 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio fre
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0719-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the provided instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM ROTATING HANDLE, W/RATCHET, INSULATED SHAFT 45CM LENGTH, P/N 0250080596 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0684-2015·2014-12-24

    Stryker Endoscopy Recalls Babcock Forceps Due to Sterilization Documentation Errors

    Stryker Endoscopy is recalling 180,573 units of Babcock Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BABCOCK FORCEPS, P/N 0250080321. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them
    Category
    Medical Device
    Distribution
    Distributed nationwide