The Recall Desk

Hazard

Loss Of Grounding recalls

2 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
2
Critical
0
Severe
0
Most recent
2018-04-04

Of the 2 loss of grounding recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all loss of grounding recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–2 of 2

  • HighFDA (Devices)·Z-1222-2018·2018-04-04

    Inova Diagnostics Recalls QUANTA-Lyser 240 EIA Due to Power Supply Defect

    Inova Diagnostics is recalling one QUANTA-Lyser 240 EIA instrument because a defective power supply may cause loss of protective grounding.

    Product
    QUANTA-Lyser 240 EIA, Part #GS0241. The instrument is a fully automated, high-throughput, bench top system designed for clinical use to automate slide and reagent barcode reading, sample dilution and distribution, reagent pipetting and dispensing, slide and plate washing, disp
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1221-2018·2018-04-04

    Inova Diagnostics Recalls QUANTA-Lyser 240 IFA Due to Power Supply Defect

    Inova Diagnostics is recalling one QUANTA-Lyser 240 IFA instrument because its power supply was not manufactured to specification, potentially causing loss of protective grounding.

    Product
    QUANTA-Lyser 240 IFA, Part #GS0242. The instrument is a fully automated, high-throughput, bench top system designed for clinical use to automate slide and reagent barcode reading, sample dilution and distribution, reagent pipetting and dispensing, slide and plate washing, disp
    Category
    Medical Device
    Distribution
    0 states