The Recall Desk

Hazard

Loose Valve Stem recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2021-03-18

Of the 1 loose valve stem recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all loose valve stem recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighNHTSA·21V193000·2021-03-18

    2021 Tiffin Phaeton recreational vehicles recalled for loose valve stem

    Tiffin is recalling certain 2021 Phaeton recreational vehicles because the valve stem stabilizer can become loose, allowing the valve stem extender to contact the wheel, which can cause tire pressure loss and increase crash risk.

    Product
    TIFFIN — TIFFIN PHAETON
    Category
    Vehicle
    Distribution
    Distributed nationwide