The Recall Desk

Hazard

Intubation Failure recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
1
Most recent
2023-11-08

Of the 1 intubation failure recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1 (100%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all intubation failure recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • SevereFDA (Devices)·Z-0144-2024·2023-11-08

    Stolen McGRATH MAC Video Laryngoscopes Recalled for Failed Quality Tests

    Covidien recalls 5,709 stolen McGRATH MAC video laryngoscopes that failed quality tests. These defective devices risk failed intubation and may compromise airway management.

    Product
    Stolen McGRATH MAC Video Laryngoscopes, REF: 301-000-000
    Category
    Medical Device
    Distribution
    0 states