The Recall Desk

Hazard

Infusion recalls

106 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
106
Critical
2
Severe
0
Most recent
2026-01-21

Of the 106 infusion recalls we have scored against our rubric, 2 (2%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all infusion recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

76–100 of 106

  • HighFDA (Devices)·Z-0607-2026·2025-12-03

    IV set 477002 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 22,928 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML, 132
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0657-2026·2025-12-03

    IV set 490427 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 224,856 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 15 DROPS/ML
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0629-2026·2025-12-03

    Extension set 456520 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 2,100 units extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. ANES SET W/DUAL FEM.LL SAFEPORT MAN
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0660-2026·2025-12-03

    IV set 490343 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 1,272 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET,2 CARESITE, TR
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0661-2026·2025-12-03

    IV set 490309 recalled over backflow into the primary bag

    B. Braun Medical is recalling 259,392 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 2 CARESITE,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0685-2026·2025-12-03

    IV set 490312 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 226,632 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMN SET W/15 DROP, ULTR
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0709-2026·2025-12-03

    Outlook set 354212 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 162,648 units outlook sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK IV SET 15DROP W/2 CARESITE; C
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0697-2026·2025-12-03

    IV set 480231 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 85,176 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0716-2026·2025-12-03

    Outlook set 354214 recalled over backflow and inability to prime

    B. Braun Medical is recalling Outlook set catalog number 354214 over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK IV SET 15DR W/2 CARESITE FILT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0663-2026·2025-12-03

    IV set 490400 recalled over backflow and inability to prime

    B. Braun Medical is recalling 19,104 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/CARESITE EXT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0641-2026·2025-12-03

    IV set 490357 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 98,688 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 10 DROPS/ML
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0651-2026·2025-12-03

    Pump administration sets recalled over piggyback backflow and priming failure

    B. Braun Medical is recalling 135,000 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET 3 CARESITE ULTRA ZER
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0675-2026·2025-12-03

    Pump IV sets recalled over backflow from piggyback into primary bags

    B. Braun Medical is recalling 65,232 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Catalog Number: 490549; Pri
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0686-2026·2025-12-03

    Low volume IV sets recalled over backflow and priming occlusion

    B. Braun Medical is recalling 960 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMINISTRATION SET, 138
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0688-2026·2025-12-03

    IV set 470179 recalled over backflow into the primary bag

    B. Braun Medical is recalling 28,344 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ B/C VALVE,1
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0700-2026·2025-12-03

    IV set 490088 recalled over medication backflow and priming failure

    B. Braun Medical is recalling iv administration sets, catalog number 490088, over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. NF1505, IV ADMIN SET W/2 H.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0683-2026·2025-12-03

    IV set 354211 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 286,400 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADM 15 DROP W/3 CARESITE
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0715-2026·2025-12-03

    Vista set US9923F recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 74,652 Vista Basic sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Vista set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. VISTA BASIC PUMP IV SET W/ 15 DROP/ML;
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0676-2026·2025-12-03

    Vented IV sets recalled over piggyback backflow and inability to prime

    B. Braun Medical is recalling 40,992 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0673-2026·2025-12-03

    IV set 490522 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 41,016 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15DR, CARESIT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0718-2026·2025-12-03

    IV set 490566 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling iv administration set catalog number 490566 over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 110 IN.; Cata
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0621-2026·2025-12-03

    Burette sets recalled over medication backflow and inability to prime

    B. Braun Medical is recalling 12,320 burette sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Burette Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. BURETTE SET, 4 CARESITE LADS, STPCK 1
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0678-2026·2025-12-03

    IV set 490542 recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 7,008 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15D, 2 CARESIT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0701-2026·2025-12-03

    IV set 490070 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 9,500 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. NF1500UP, IV ADMINISTRATION
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0624-2026·2025-12-03

    Caresite set 354204 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 12,225,913 units caresite sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. Primary Gravity IV Set with 2 CARESITE¿
    Category
    Medical Device
    Distribution
    0 states