REMISOL Advance Medical Device Software Defect Affecting Reference Range Calculations
A software issue in REMISOL Advance may incorrectly process date-of-birth information from laboratory systems, causing incorrect reference range calculations and potentially auto-validating erroneous test result flags. The actual test results are not changed, but their evaluation may be affected.
- Product
- REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164; C44703; C57017; C69412; C69413; C73942; C73941; C88470; C88471 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 con
- Category
- Medical Device
- Distribution
- Distributed nationwide