The Recall Desk

Hazard

Incorrect Diagnosis Risk recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2026-02-11

Of the 1 incorrect diagnosis risk recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all incorrect diagnosis risk recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighFDA (Devices)·Z-1235-2026·2026-02-11

    Sample preparation systems recalled over a software dispensing error

    Beckman Coulter is recalling 50 CellMek SPS Sample Preparation Systems because a software error dispenses samples close to or touching tube bottoms, removing 60-100 microliters from the final sample.

    Product
    CellMek SPS Sample Preparation System, REF: C44603, with software
    Category
    Medical Device
    Distribution
    Distributed nationwide