The Recall Desk

Hazard

Impedance Rise recalls

3 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
3
Critical
3
Severe
0
Most recent
2025-08-27

Of the 3 impedance rise recalls we have scored against our rubric, 3 (100%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all impedance rise recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–3 of 3

  • CriticalFDA (Devices)·Z-2327-2025·2025-08-27

    RELIANCE 4-FRONT defibrillation leads recalled over rising LVSI measurements and coil calcification

    RELIANCE 4-FRONT defibrillation leads are recalled because affected models may show gradually rising LVSI measurements linked to calcification of the defibrillation coils, which may lead to early lead replacement.

    Product
    RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • CriticalFDA (Devices)·Z-2325-2025·2025-08-27

    ENDOTAK RELIANCE defibrillation leads recalled over coil calcification and rising LVSI

    ENDOTAK RELIANCE G/SG transvenous defibrillation leads are recalled because affected models may show gradually rising LVSI measurements linked to calcification of the defibrillation coils.

    Product
    ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • CriticalFDA (Devices)·Z-2326-2025·2025-08-27

    ENDOTAK RELIANCE leads with 4-SITE connector recalled over coil calcification

    ENDOTAK RELIANCE G/SG defibrillation leads with the 4-SITE connector are recalled because affected models may show gradually rising LVSI measurements linked to calcification of the defibrillation coils.

    Product
    ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296
    Category
    Medical Device
    Distribution
    Distributed nationwide