7 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
7
Critical
0
Severe
0
Most recent
2026-04-22

Of the 7 imaging system recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all imaging system recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–7 of 7

  • HighFDA (Devices)·Z-1825-2026·2026-04-22

    Philips Spectral CT on Rails recalled over three software issues

    Philips is recalling the Spectral CT on Rails after identifying three software issues affecting gantry movement, a confirmation prompt and head area planning. Three systems are affected.

    Product
    Philips Spectral CT on Rails. Model Number: 728334.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1814-2026·2026-04-22

    ARTIS Pheno 3D acquisitions may capture uneven patient image brightness

    ARTIS Pheno VE30A and VE40A systems are being recalled because 3D acquisitions may capture lighter and darker images, reducing reconstruction accuracy. 73 units are affected.

    Product
    ARTIS Pheno VE30A and VE40A, Model 10849000
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-1653-2026·2026-04-15

    Siemens ARTIS imaging systems may display an inflated patient dose reading

    Siemens ARTIS pheno and ARTIS Icono systems may sporadically display a major increase in the x-ray dose applied, although the actual dose applied remains within system settings and regulatory limits.

    Product
    ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2502-2025·2025-09-10

    Azurion imaging systems recalled because single phase UPS may fail on power loss

    Philips Azurion systems configured with a 1 Phase UPS are recalled because they may shut down and/or fail to start up if power to the system is lost. 1,861 units are affected.

    Product
    Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Azurion 3 M12, Model Number 722063. 2. Azurion 3 M15, Model Number 722064. 3. Azurion 7 B12, Model Number 722067. 4. Az
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2503-2025·2025-09-10

    Interventional imaging systems recalled because single phase UPS may fail on power loss

    Philips Allura systems configured with a 1 Phase UPS are recalled because they may shut down and/or fail to start up if power to the system is lost. 1,006 units are affected.

    Product
    Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2504-2025·2025-09-10

    Fluoroscopic X-ray systems recalled over air baffle missing from internal PCs

    Philips Allura Xper interventional fluoroscopic X-ray systems are recalled because an air baffle component may be missing in some Nehalem FlexVision-2 PCs. 192 units are affected.

    Product
    Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003. 2. Allura Xper FD10/10, Model Number 722005. 3. Allura Xper FD20, Model Number 722006. 4. All
    Category
    Medical Device
    Distribution
    Distributed nationwide