Siemens Artis Q Biplane X-ray System Software Imaging Delay Issue
Siemens Medical Solutions is recalling 15 Artis Q biplane diagnostic x-ray systems due to a software issue in the VD12-angiography subsystem. When source-to-image distance lift is activated simultaneously with x-ray release, the system fails to activate and displays an error message, causing procedural delay.
- Product
- Artis Q biplane with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848282
- Category
- Medical Device
- Distribution
- Distributed nationwide