The Recall Desk

Hazard

Imaging Degradation recalls

2 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
2
Critical
0
Severe
0
Most recent
2024-04-03

Of the 2 imaging degradation recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all imaging degradation recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–2 of 2

  • HighFDA (Devices)·Z-1396-2024·2024-04-03

    Certain OptiCross Coronary Imaging Catheters recalled for manufacturing defect

    Boston Scientific recalls certain batches of OptiCross Coronary Imaging Catheters manufactured with an incorrect voltage setting that may cause decreased image brightness.

    Product
    OptiCross Coronary Imaging Catheter, REF H749518080120
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2474-2021·2021-09-22

    Philips Allura Xper X-ray imaging system failures at 480V power supply

    Four Philips Allura Xper X-ray imaging systems may fail prematurely when powered at 480V hospital mains, causing coil overheating and reduced imaging performance.

    Product
    The Allura Xper series is intended for use on human patients to perform: "Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal a
    Category
    Medical Device
    Distribution
    Distributed nationwide