The Recall Desk

Hazard

Glucose Reading Failure recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2025-06-11

Of the 1 glucose reading failure recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all glucose reading failure recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighFDA (Devices)·Z-1906-2025·2025-06-11

    Dexcom G7 iOS App version 2.7.0 stops updating glucose readings

    Dexcom G7 iOS App version 2.7.0 can enter a state where glucose readings freeze and alerts do not trigger, risking missed detection of dangerously high or low blood sugar. Other connected devices like Apple Watch and insulin pumps continue to receive glucose data directly from the transmitter.

    Product
    Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system
    Category
    Medical Device
    Distribution
    Distributed nationwide