The Recall Desk

Hazard

Fall Risk recalls

118 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
118
Critical
1
Severe
13
Most recent
2026-08-12

Of the 118 fall risk recalls we have scored against our rubric, 1 (1%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 13 (11%) are Severe.

Which agencies issued them

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all fall risk recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

26–50 of 118

  • HighFDA (Devices)·Z-1595-2024·2024-04-24

    SW Bari Lift & Transfer Device Recalled Due to Weight and Safety Defects

    Raye's Inc. is recalling the SW Bari Lift & Transfer (Model 38060000) due to two critical defects: it cannot safely support its maximum rated weight, and it lacks a safety mechanism to prevent falls during a single equipment failure.

    Product
    SW Bari Lift & Transfer, Model Number 38060000
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1574-2024·2024-04-24

    MRI patient support table floor plate may be incorrectly installed

    Philips Intera 1.0T MRI systems may have patient support table floor plates that are incorrectly installed. The defect poses a potential safety risk to patients. Facilities should verify installation and contact Philips service personnel for inspection.

    Product
    Intera 1.0T Power/Pulsar, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1575-2024·2024-04-24

    Philips MRI Patient Support Table Floor Plate May Be Incorrectly Installed

    Philips is recalling 347 Intera 1.5T MRI systems worldwide because the patient support table floor plate may be incorrectly installed, potentially compromising patient support stability during use.

    Product
    Intera 1.5T, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1581-2024·2024-04-24

    Philips Intera 1.5T MRI System Patient Support Table Installation Defect

    Philips is recalling certain Intera 1.5T MRI systems because the patient support table floor plate may be incorrectly installed, which could affect patient safety.

    Product
    Intera 1.5T Omni/Stellar, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1567-2024·2024-04-24

    MRI System Patient Support Table Floor Plate Installation Defect

    Philips Ingenia 3.0T CX MRI systems may have an incorrectly installed patient support table floor plate. This installation error could compromise the stability of the patient support table during use.

    Product
    Ingenia 3.0T CX, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1583-2024·2024-04-24

    Philips MRI System Patient Support Table Installation Defect Recall

    Philips recalls 25 Intera 1.5T R11 MRI systems due to potential incorrect installation of the patient support table floor plate, which may affect patient safety during device operation.

    Product
    Intera 1.5T R11, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1506-2024·2024-04-17

    Chromophare Surgical Light System recalled for insufficient mounting force

    Stryker Communications recalls Chromophare Surgical Light Systems due to insufficient mounting force that could compromise the equipment's stability in surgical environments.

    Product
    Chromophare Surgical Light System , REF CH00000001
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1487-2024·2024-04-17

    Medline Adult Portable Bed Rails Recalled for Safety Standard Failures

    Medline is recalling certain lots of adult portable bed rails that were manufactured after the mandatory compliance date but failed to meet required safety standards, potentially posing a fall risk.

    Product
    Bed Assist Bars, labeled as follows: a) Medline, Model Number BA6800, b) Medline Martha Stewart, Model Number MDS6800BAMMS
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1486-2024·2024-04-17

    Patient lift hanger bar may come loose, posing fall risks

    Certain patient lift models may have a hanger bar that could come loose during use, potentially causing patient falls. Approximately 10 units distributed to TX, IL, OK, and CA are affected.

    Product
    Altair 330 lbs w/o Hanger Bar, REF: 55200H-NA, 55220H-NA, and Roomer S 220 kg w/o Hanger Bar, REF: 55300H-NA, patient lifts
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0959-2024·2024-02-21

    GE Healthcare Care Plus Incubator Hood Latching Defect Increases Fall Risk

    GE Healthcare incubator hoods can appear closed but may not be properly latched, posing a fall risk if patients impact an unsecured panel.

    Product
    GE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 6600-0047-851, 6600-0033-850, 6600-0081-850, 6600-0961-500, 6600-0083-850
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0958-2024·2024-02-21

    GE Healthcare Care Plus and Lullaby incubators recalled for bedside panel latching defect

    GE Healthcare is recalling 1,410 neonatal incubators worldwide because bedside panels and portholes can appear closed while not properly latched. If impacted, they may fall open and fail to protect patients from falls.

    Product
    GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby Incubator TR
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0957-2024·2024-02-21

    GE Healthcare neonatal incubator bedside panels may fail to latch properly

    GE Healthcare is recalling neonatal incubators because the bedside panel can appear closed while unlatched. Impact could cause it to fall open, risking patient injury.

    Product
    GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000, c) Care Plus 2000, d) Care Plus 3000, e) Care Plus 4000, Plus 4000
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0961-2024·2024-02-21

    GE Healthcare Incubators Bedside Panel Latch Failure Poses Fall Risk

    GE Healthcare incubators have defective bedside panels that may appear closed but not be properly latched, creating a patient fall risk when the panels disengage upon impact.

    Product
    GE Healthcare Lullaby Incubator Model Hood FRU kit, Part Numbers: M1169566, M1160659, M1206259, M1206260, M1166836, M1166814
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0960-2024·2024-02-21

    GE Healthcare incubator bedside panel may disengage and fall open

    The bedside panel on GE Healthcare Care Plus incubators may appear closed while unlatched. If impacted, the panel can fall open, exposing patients to injury risk.

    Product
    GE Healthcare Care Plus Hood Door FRU kit, Part Numbers: 6600-0082-850, 6600-0504-800, 6600-0148-800, and 6600-0970-500
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0788-2024·2024-01-31

    Power Knee Prosthetic Batteries Recalled Due to Unexpected Shutdown Risk

    Ossur Power Knee prosthetic knee batteries may fail to turn on or shut down unexpectedly, potentially causing users to lose support and fall. Affected units are distributed nationwide.

    Product
    Power Knee, REF: PKA10001, and PKA10003; with Power Knee Battery, REF: PKA10002
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0762-2024·2024-01-24

    Maquet Orchide OR Light Systems recalled due to fall risk

    Getinge USA is recalling Maquet Orchide OR Light Systems due to a potential for light fixtures to fall in the operating room. No injuries have been reported.

    Product
    Maquet Orchide OR Light Systems, Model Numbers and Catalog Numbers OHDII FHD QL VP01 - ARD568803935, OHDII FHD QL AIR03 U - ARD568803952, OHDII FHD QL AIR03 E - ARD568803967, OHDII FHD QL AIR04 E/U - ARD568803968, OHDII FHD QL+ VP01 - ARD569204944, OHDII AIR03 QL FHD E - ARD56880
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0769-2024·2024-01-24

    Maquet G8 Operating Room Light Systems Recalled for Fall Risk

    Getinge USA is recalling Maquet G8 and G8E operating room light systems that may fall during use. Approximately 236,793 units have been distributed nationwide and globally.

    Product
    Maquet G8 / G8E OR Light Systems, Model Numbers and Catalog Numbers G8 LC E - ARD568131211C, G8G8 DF E - ARD568121111C, G8G8FS DF - ARD568121131C
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0760-2024·2024-01-24

    Maquet Rolite OR Light Systems Recall Due to Fall Risk

    Getinge USA is recalling 236,793 Maquet Rolite operating room light systems due to a potential for the light fixture to fall. No injuries have been reported.

    Product
    Maquet Rolite OR Light Systems, Model Numbers and Catalog Numbers ROL PWDII50SF - ARDPWT309001A, ROL PWDII50SF - ARDPWT309002A, ROL PWDII50SF - ARDPWT309009A, ROL PWDII50SF - ARDPWT309010A, ROL PWDII70SF - ARDPWT309007A, ROL PWDII70SF - ARDPWT309008A, ROL PWDII70SF - ARDPWT309015
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0231-2024·2023-11-15

    Breg Folding Walkers With Wheels Recalled for Misaligned Wheels

    Breg Inc is recalling 11,193 folding walkers with misaligned wheels that do not meet product specifications. The defect could affect stability and increase fall risk.

    Product
    BREG REF 100519-000 WALKER FOLDING WALKER WITH WHEELS Utilized as a walking aid.
    Category
    Medical Device
    Distribution
    41 states
  • HighFDA (Devices)·Z-2052-2023·2023-07-12

    Treadmill Recalled for Drive PCB Malfunction Causing Unexpected Belt Movement

    Full Vision Inc is recalling approximately 565 TrackMaster TMX428 treadmills due to a drive circuit board malfunction that can cause the belt to unexpectedly stop, reverse, or accelerate, posing a risk of falls or injury.

    Product
    TRACKMASTER, TREADMILL TMX428 110V, MODEL NUMBER 317-07926
    Category
    Consumer Product
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2053-2023·2023-07-12

    TrackMaster TMX428 Treadmill drive circuit board malfunction recall

    Drive board malfunction in TrackMaster TMX428 treadmills can cause unexpected belt reversal or acceleration, creating fall and impact injury risks. 376 units affected.

    Product
    TRACKMASTER, TREADMILL TMX428 220V, MODEL NUMBER 317-07927
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2058-2023·2023-07-12

    Treadmill recalled for drive PCB malfunction causing unexpected acceleration

    Full Vision Inc recalls 95 TRACKMASTER GE T2100-ST2 treadmills with a defective drive PCB that may misread signals, causing the equipment to unexpectedly stop, reverse, or accelerate forward.

    Product
    TRACKMASTER, TREADMILL GE T2100-ST2 220V CHINA, MODEL NUMBER 317-07927GE CHINA
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1000-2023·2023-01-25

    Medical X-ray equipment patient table footrests recalled due to detachment risk

    Siemens recalls four fluoroscopy and radiographic imaging equipment models worldwide because patient table footrests may detach during clinical use, potentially causing patient injury.

    Product
    AXIOM Luminos TF, AXIOM Iconos R100, AXIOM Iconos R200 and AXIOM Luminos dRF
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1001-2023·2023-01-25

    Patient table footrest may detach from AXIOM Luminos Agile

    Siemens Medical Solutions is recalling the AXIOM Luminos Agile patient table because the footrest may detach during use, creating a potential injury risk.

    Product
    AXIOM Luminos Agile
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1002-2023·2023-01-25

    X-ray imaging system footrests may detach from patient tables

    Siemens Luminos X-ray imaging systems may have a footrest that detaches from the patient table during use. The manufacturer is recalling 886 units distributed worldwide.

    Product
    Luminos Agile Max, Luminos dRF Max, and LUMINOS Lotus Max
    Category
    Medical Device
    Distribution
    0 states