The Recall Desk

Hazard

Elevated Iop recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2026-03-11

Of the 1 elevated iop recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all elevated iop recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighFDA (Devices)·Z-1480-2026·2026-03-11

    Implantable collamer lenses recalled after being labeled with wrong length

    Staar Surgical is recalling seven EVO+ Visian Toric Implantable Collamer Lenses that were labeled as 12.1 mm but are 12.6 mm, creating the potential for excessive vault.

    Product
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
    Category
    Medical Device
    Distribution
    0 states