Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all dialysate recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighFDA (Devices)·Z-1185-2026·2026-02-04

    Conductivity standard solutions recalled over incorrect dialysate conductivity values

    NIPRO Technical Services is recalling 499 units of Conductivity Standard Solution because a defect in the solution can cause meters to report incorrect conductivity values for dialysate.

    Product
    Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters
    Category
    Medical Device
    Distribution
    Distributed nationwide