The Recall Desk

Hazard

Contrast Overdose recalls

3 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
3
Critical
0
Severe
0
Most recent
2021-08-25

Of the 3 contrast overdose recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all contrast overdose recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–3 of 3

  • HighFDA (Devices)·Z-2317-2021·2021-08-25

    IQon Spectral CT System Software Error Increases Contrast Volume

    Philips IQon Spectral CT systems can recommend excessive contrast volume when patient weight is set to pounds due to a SynchRight software calculation error, potentially resulting in patient overdose.

    Product
    728332 IQon Spectral CT-Computed Tomography X-ray system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2315-2021·2021-08-25

    CT5000 X-ray System Contrast Dose Calculation Error Under Pounds Setting

    Philips CT5000 X-ray systems may incorrectly calculate contrast volume doses when patient weight units are set to pounds. The software error could recommend higher-than-intended contrast volumes.

    Product
    Model 728321/728323 CT5000 Ingenuity Plus/Pro/Premium-Computed Tomography X-ray system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2316-2021·2021-08-25

    CT system software error recommends excessive contrast volume

    A software defect in Philips CT imaging systems causes incorrect contrast volume calculations when patient weight units are set to pounds, resulting in increased contrast dosing recommendations. Three units with specific serial numbers are affected.

    Product
    728327 Ingenuity CT Upgrades-Computed Tomography X-ray system
    Category
    Medical Device
    Distribution