The Recall Desk

Hazard

Connector Movement recalls

2 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
2
Critical
0
Severe
0
Most recent
2022-07-27

Of the 2 connector movement recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all connector movement recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–2 of 2

  • HighFDA (Devices)·Z-1448-2022·2022-07-27

    Cervical spinal implant connectors may fail to maintain proper clamping force

    Aesculap Implant Systems has recalled certain cervical spine implant connectors due to insufficient clamping force. The defect may allow movement on the rods or cause microparticle abrasion.

    Product
    AESCULAP S4 Cervical Cross Connector L 42-58mm, S4C CROSS CONNECTOR VARIABLE 42-58MM, product code SW117T. Used in spinal fusion surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1444-2022·2022-07-27

    Aesculap S4 Cervical Cross Connector Spinal Implant System Recalled for Clamping Failure

    Aesculap Implant Systems is recalling the AESCULAP S4 Cervical Cross Connector (product code SWT113T) used in spinal fusion surgery. The connectors may not clamp properly onto rods, potentially causing movement and wear particles.

    Product
    AESCULAP S4 Cervical Cross Connector L24mm, S4C CROSS CONNECTOR FIXED 24MM, product code SWT113T. Used in spinal fusion surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide