The Recall Desk

Hazard

Cold Weather recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2025-02-13

Of the 1 cold weather recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all cold weather recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighNHTSA·25V085000·2025-02-13

    Rivian R1S and R1T low beams may fail in cold weather

    Rivian is recalling certain 2025 R1S and R1T vehicles because the headlight low beams may fail to illuminate when starting the vehicle in cold weather, failing FMVSS 108.

    Product
    RIVIAN — RIVIAN R1S, redundant R 1 T
    Category
    Vehicle
    Distribution
    Distributed nationwide