The Recall Desk

Hazard

Carbon Monoxide Inhalation recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2006-04-20

Of the 1 carbon monoxide inhalation recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all carbon monoxide inhalation recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighCPSC·06144·2006-04-20

    Emergency Smoke Hoods Recalled for Carbon Monoxide Risk

    CPSC recalls EVAC-U8 and EVAC+ smoke hoods because they may fail to filter carbon monoxide. Consumers should stop using and dispose of the devices.

    Product
    EVAC-U8™ and EVAC+™ Emergency Escape Smoke Hoods
    Category
    Consumer Product
    Distribution
    Distributed nationwide